A FORCE TRANSDUCER AND A LENGTH-RAMP GENERATOR FOR MECHANICAL INVESTIGATIONS OF FROG-HEART MYOCYTES

被引:15
作者
CECCHI, G [1 ]
COLOMO, F [1 ]
POGGESI, C [1 ]
TESI, C [1 ]
机构
[1] UNIV FLORENCE, DIPARTIMENTO SCI FISIOL, VIALE G B MORGAGNI 63, I-50134 FLORENCE, ITALY
来源
PFLUGERS ARCHIV-EUROPEAN JOURNAL OF PHYSIOLOGY | 1993年 / 423卷 / 1-2期
关键词
FROG-HEART MYOCYTES; FORCE TRANSDUCER; LENGTH-RAMP GENERATOR;
D O I
10.1007/BF00374968
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
An apparatus for studying the mechanics of isolated frog heart myocytes is described. The cells are held horizontal in a through of Ringer solution by means of two suction micropipettes. Myocyte force is measured with an opto-electronic system recording the deflection of the tip of one micropipette, which acts as a cantilever force probe. The force probes are selected for compliance according to the force a myocyte is expected to develop in a given condition, so as to limit myocyte shortening during force development to no more than 1% of the slack cellular length (l0). The other micropipette, which is stiff relative to the forces measured, is mounted on an electromagnetic-loudspeaker motor by which controlled-velocity length changes, of preset size and in either direction, are imposed on myocytes. The force transducer has a sensitivity of 5-10 mV/nN, with a frequency response of 700-900 Hz in Ringer solution and a resolution of 0.5-1 nN. The motor with a suction micropipette can complete controlled-velocity length ramps within 1.5-2.0 ms, across a range of +/- 100 mum at a resolution of 8.0 nm. These values correspond, for frog-heart myocytes 200 mum and 400 mum long, to 25%-50% l0 and 0.002%-0.004% l0 respectively.
引用
收藏
页码:113 / 120
页数:8
相关论文
共 23 条