FLUOROSCOPIC IMAGE AS INPUT DATA FOR COMPUTED-TOMOGRAPHY

被引:4
作者
BAILY, NA
机构
[1] Department of Radiology, University of California
[2] La Jolla, California 92093
关键词
D O I
10.1109/TNS.1979.4330518
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The use of a fluoroscopic imaging system for providing the projection data required in computerized tomography has been explored. The capability of both conventional systems and large flat screen-low light level video systems are discussed. The highlights of previously published data are also presented. Specific applications to radiation therapy are illustrated and the equipment described. The differential contrast sensitivity is of the order of 0.5%, spatial resolution less than 0.5 mm, and slice thickness approximately 1mm. Patient skin exposure is expected to be between 150 mR and 300 mR per scan. © 1979 IEEE.