摘要
The shortcomings of neutron remmeters have been well documented by various research laboratories and discussed by the ICRU. Nonetheless, due to a forced reliance on them at the operational level there is frequently an undue confidence in them. This paper reviews the concept of dose equivalent index and how accurately neutron remmeters measure this quantity. It is shown that both instrumental and conceptual shortcomings of mean remmeters can overestimate the dose equivalent index by up to a factor of 15 in the worst case, and up to a factor of 3 or 4 in situations which are not unusual.
引用
收藏
页码:735 / 742
页数:8
相关论文
共 22 条
  • [1] Andersson I.O., 1963, NEUTRON DOSIMETRY, VII, P87
  • [2] BARTINE DE, 1977, ORNLTM4840 REP
  • [3] FACEY RA, 1978, 1978 HLTH PHYS SOC M
  • [4] HANKINS DE, 1975, HEALTH PHYS, V28, P305
  • [5] MODIFIED A-B REMMETER WITH IMPROVED DIRECTIONAL DEPENDENCE AND THERMAL-NEUTRON SENSITIVITY
    HANKINS, DE
    [J]. HEALTH PHYSICS, 1978, 34 (03): : 249 - 254
  • [6] HANKINS DE, 1977, 6TH ERDA WORKSH PERS
  • [7] HANKINS DE, UNPUBLISHED
  • [8] APPROACH TO SURVEYING OF RADIATION ENVIRONMENTS FOR RADIATION PROTECTION PURPOSES
    HARVEY, JR
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1975, 20 (06) : 1003 - 1014
  • [9] CALIBRATION OF A NUMBER OF NEUTRON SENSITIVE INSTRUMENTS WITH A NOVEL SOURCE EMITTING NEUTRONS AROUND O.5 KEV
    HARVEY, JR
    LAVENDER, A
    THOMPSON, IMG
    [J]. HEALTH PHYSICS, 1976, 31 (04): : 363 - 369
  • [10] KLUGE H, 1972, NEUTRON MONITORING R, V1, P13