COMPARATIVE VALIDATION OF QUANTITATIVE CORONARY ANGIOGRAPHY SYSTEMS - RESULTS AND IMPLICATIONS FROM A MULTICENTER STUDY USING A STANDARDIZED APPROACH

被引:141
作者
KEANE, D
HAASE, J
SLAGER, CJ
VANSWIJNDREGT, EM
LEHMANN, KG
OZAKI, Y
DIMARIO, C
KIRKEEIDE, R
SERRUYS, PW
机构
[1] ERASMUS UNIV ROTTERDAM,THORAXCTR,DEPT INTERVENT CARDIOL,CARDIAC CATHETERIZAT LAB,3000 DR ROTTERDAM,NETHERLANDS
[2] ERASMUS UNIV ROTTERDAM,THORAXCTR,DEPT INTERVENT CARDIOL,INTRACORONARY IMAGING LAB,3000 DR ROTTERDAM,NETHERLANDS
[3] ERASMUS UNIV ROTTERDAM,THORAXCTR,DEPT INTERVENT CARDIOL,EXPTL CARDIOL LAB,3000 DR ROTTERDAM,NETHERLANDS
[4] UNIV TEXAS,HLTH SCI CTR,HOUSTON,TX
关键词
ANGIOGRAPHY; CORONARY DISEASE; STENOSIS;
D O I
10.1161/01.CIR.91.8.2174
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Computerized quantitative coronary angiography (QCA) has fundamentally altered our approach to the assessment of coronary interventional techniques and strategies aimed at the prevention of recurrence and progression of stenosis. It is essential, therefore, that the performance of QCA systems, upon which much of our scientific understanding has become integrally dependent, is evaluated in an objective and uniform manner. Methods and Results We validated 10 QCA systems at core laboratories in North America and Europe. Cine films were made of phantom stenoses of known diameter (0.5 to 1.9 mm) under four experimental conditions: in vivo (coronary arteries of pigs) calibrated at the isocenter or by use of the catheter as a scaling device and in vitro with 50% contrast and 100% contrast. The cine films were analyzed by each automated QCA system without observer interaction. Accuracy and precision were taken as the mean and SD of the signed differences between the phantom stenoses, and the measured minimal luminal diameters and the correlation coefficient (r), the SEE, the y intercept, and the slope were derived by their linear regression. Performance of the 10 QCA systems ranged widely: accuracy, +0.07 to +0.31 mm; precision, +/-0.14 to +/-0.24 mm; correlation (r), .96 to .89; SEE, +/-0.11 to +/-0.16 mm; intercept, +0.08 to +0.31 mm; and slope, 0.86 to 0.64. Conclusions There is a marked variability in performance between systems when assessed over the range of 0.5 to 1.9 mm. The range of accuracy, intercept, and slope values of this report indicates that absolute measurements of luminal diameter from different multicenter angiographic trials may not be directly comparable and additionally suggests that such absolute measurements may not be directly applicable to clinical practice using an on-line QCA system with a different edge detection algorithm. Power calculations and study design of angiographic trials should be adjusted for the precision of the QCA system used to avoid the risk of failing to detect small differences in patient populations. This study may guide the fine-tuning of algorithms incorporated within each system and facilitate the maintenance of high standards of QCA for scientific studies.
引用
收藏
页码:2174 / 2183
页数:10
相关论文
共 59 条
[1]   A COMPARISON OF DIRECTIONAL ATHERECTOMY WITH BALLOON ANGIOPLASTY FOR LESIONS OF THE LEFT ANTERIOR DESCENDING CORONARY-ARTERY [J].
ADELMAN, AG ;
COHEN, EA ;
KIMBALL, BP ;
BONAN, R ;
RICCI, DR ;
WEBB, JG ;
LARAMEE, L ;
BARBEAU, G ;
TRABOULSI, M ;
CORBETT, BN ;
SCHWARTZ, L ;
LOGAN, AG .
NEW ENGLAND JOURNAL OF MEDICINE, 1993, 329 (04) :228-233
[2]   CHANGE IN DIAMETER OF CORONARY-ARTERY SEGMENTS ADJACENT TO STENOSIS AFTER PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY - FAILURE OF PERCENT DIAMETER STENOSIS MEASUREMENT TO REFLECT MORPHOLOGIC CHANGES INDUCED BY BALLOON DILATION [J].
BEATT, KJ ;
LUIJTEN, HE ;
DEFEYTER, PJ ;
VANDENBRAND, M ;
REIBER, JHC ;
SERRUYS, PW ;
TENKATEN, HJ ;
VANES, GA .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1988, 12 (02) :315-323
[3]   RESTENOSIS AFTER CORONARY ANGIOPLASTY - THE PARADOX OF INCREASED LUMEN DIAMETER AND RESTENOSIS [J].
BEATT, KJ ;
SERRUYS, PW ;
LUIJTEN, HE ;
RENSING, BJ ;
SURYAPRANATA, H ;
DEFEYTER, P ;
VANDENBRAND, M ;
LAARMAN, GJ ;
ROELANDT, J ;
VANES, GA .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 19 (02) :258-266
[4]   RESTENOSIS AFTER CORONARY ANGIOPLASTY - NEW STANDARDS FOR CLINICAL-STUDIES [J].
BEATT, KJ ;
SERRUYS, PW ;
HUGENHOLTZ, PG .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 15 (02) :491-498
[5]  
BEIER J, 1992, COMPUT CARDIOL, P513
[6]   QUANTITATIVE CORONARY ANGIOGRAPHY TO MEASURE PROGRESSION AND REGRESSION OF CORONARY ATHEROSCLEROSIS - VALUE, LIMITATIONS, AND IMPLICATIONS FOR CLINICAL-TRIALS [J].
DEFEYTER, PJ ;
SERRUYS, PW ;
DAVIES, MJ ;
RICHARDSON, P ;
LUBSEN, J ;
OLIVER, MF .
CIRCULATION, 1991, 84 (01) :412-423
[7]   MATCHING BASED ON QUANTITATIVE CORONARY ANGIOGRAPHY AS A SURROGATE FOR RANDOMIZED STUDIES - COMPARISON BETWEEN STENT IMPLANTATION AND BALLOON ANGIOPLASTY OF NATIVE CORONARY-ARTERY LESIONS [J].
DEJAEGERE, PP ;
HERMANS, WR ;
RENSING, BJ ;
STRAUSS, BH ;
DEFEYTER, PJ ;
SERRUYS, PW .
AMERICAN HEART JOURNAL, 1993, 125 (02) :310-319
[8]   VARIABILITY IN ANALYSIS OF CORONARY ARTERIOGRAMS [J].
DEROUEN, TA ;
MURRAY, JA ;
OWEN, W .
CIRCULATION, 1977, 55 (02) :324-328
[9]   OBSERVER AGREEMENT IN EVALUATING CORONARY ANGIOGRAMS [J].
DETRE, KM ;
WRIGHT, E ;
MURPHY, ML ;
TAKARO, T .
CIRCULATION, 1975, 52 (06) :979-986
[10]  
DIMARIO C, 1992, AM HEART J, V124, P1181