FAST SPIN-ECHO STUDIES OF CONTRAST AND SMALL-LESION DEFINITION IN A LIVER-METASTASIS PHANTOM

被引:15
作者
CHIEN, D [1 ]
MULKERN, RV [1 ]
机构
[1] VET ADM MED CTR,DEPT RADIOL,SAN FRANCISCO,CA 94121
来源
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING | 1992年 / 2卷 / 04期
关键词
LIVER NEOPLASMS; PHANTOMS; PHASE-ENCODING ORDER; PULSE SEQUENCES; RAPID IMAGING;
D O I
10.1002/jmri.1880020419
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A liver-metastasis model was used to study the ability of fast spin-echo (FSE) imaging to show small lesions (1 pixel in diameter) relative to conventional spin-echo imaging. FSE images of the liver-metastasis phantom were acquired with various phase-encode reordering schemes to manipulate T2 contrast. The imaging time for multisection acquisitions was 27 seconds for FSE imaging and 6 minutes 48 seconds for conventional spin-echo imaging. Computer simulations were performed to determine how the point spread function varies with the different phase-encoding orders in FSE imaging. Contrast-to-noise ratios and signal profiles of the lesions were measured as a function of the effective TE and lesion size. Experimental results and theoretical simulations showed that T2-weighted FSE imaging provides high contrast and good edge definition even for small lesions. The results indicate that FSE imaging may become a powerful method for the early detection of liver metastases.
引用
收藏
页码:483 / 487
页数:5
相关论文
共 12 条
  • [1] Hennig J, Naureth A, Friedburg H, RARE imaging: a fast imaging method for clinical MR, Magn Reson Med, 3, pp. 823-833, (1986)
  • [2] Mulkern RV, Wong STS, Winalski C, Jolesz FA, Contrast manipulation and artifact assessment of 2D and 3D RARE sequences, Magn Reson Imaging, 8, pp. 557-566, (1990)
  • [3] Melki PS, Mulkern RV, Panych LP, Jolesz FA, Comparing the FAISE method with conventional dual‐echo sequences, JMRI, 1, pp. 319-326, (1991)
  • [4] Mulkern RV, Melki PS, Jakab P, Higuchi N, Jolesz FA, Phase encode order and its effect on contrast and artifact in single‐shot RARE sequences, Med Phys, 18, pp. 1032-1037, (1991)
  • [5] Jones KM, Mulkern RV, Mantello MT, Et al., Brain hemorrhage: evaluation with fast spin‐echo and conventional dual spin‐echo images, Radiology, 182, pp. 53-58, (1992)
  • [6] Melboom S, Gill D, Modified spin‐echo method for measuring nuclear relaxation times, Review of Scientific Instruments, 29, pp. 688-691, (1958)
  • [7] Melki PS, Mulkern RV, Jolesz FA, How fast are fast spin‐echo techniques compared with spin‐echo sequences? (abstr), Radiology, 181, P, (1991)
  • [8] Chien D, Atkinson DJ, Edelman RR. Strategies to improve contrast in turbo‐FLASH imaging: reordered phase encoding and k‐space segmentation, JMRI, 1, pp. 63-70, (1991)
  • [9] Norris DG, Bottcher U, Leibfritz D, A simple method of generating variable T1 contrast using temporally reordered phase encoding, Magn Reson Med, 15, pp. 483-490, (1990)
  • [10] Higuchi N, Oshio K, Imai Y, Et al., Clinical applications of multishot RARE in abdominal MR imaging (abstr), JMRI, 1, pp. 150-151, (1991)