PHYSICS OF GAMMA-KNIFE APPROACH ON CONVERGENT BEAMS IN STEREOTAXIC RADIOSURGERY

被引:208
作者
WU, A
LINDNER, G
MAITZ, AH
KALEND, AM
LUNSFORD, LD
FLICKINGER, JC
BLOOMER, WD
机构
[1] UNIV PITTSBURGH,SCH MED,JOINT RADIAT ONCOL CTR,DEPT RADIAT ONCOL,PITTSBURGH,PA 15261
[2] UNIV PITTSBURGH,SCH MED,JOINT RADIAT ONCOL CTR,DEPT NEUROSURG,PITTSBURGH,PA 15261
[3] PITTSBURGH CANC INST,PITTSBURGH,PA 15213
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1990年 / 18卷 / 04期
关键词
'Co; Dosimetry; Radiosurgery; Stereotactic frame;
D O I
10.1016/0360-3016(90)90421-F
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
The Presbyterian-University Hospital of Pittsburgh installed the first clinically designated Leksell gamma knife in the U.S. in August 1987. Gamma knife radiosurgery involves stereotactic target localization with the Leksell frame and subsequent closed-skull single-treatment session irradiation of a lesion with multiple highly focused gamma ray beams produced from 60Co sources. The hemispherical array of sources, the large number of small-diameter beams, and the steep dose gradients surrounding a targeted lesion make physical characterization of the radiation field complex. This paper describes the physical features and the operation of the gamma knife as well as the calibration procedures of the very small, well-collimated beams. The results of studies using in-phantom ion chamber, diode, film, and lithium fluoride thermoluminescent dosimetry were all in close agreement. Both single-beam and multiple-beam dose profiles were measured and reported for the interchangeabale helmets, which have 4-, 8-, 14-, and 18-mm-diameter collimators. We also describe the dose calculation and treatment planning algorithm in the treatment planning system. Measurements of the accuracy of mechanical and radiation alignment are also performed and discussed. © 1990.
引用
收藏
页码:941 / 949
页数:9
相关论文
共 27 条
[1]   THE HUMAN ACOUSTIC NEURINOMA IN ORGAN-CULTURE .2. TISSUE CHANGES AFTER GAMMA-IRRADIATION [J].
ANNIKO, M ;
ARNDT, J ;
NOREN, G .
ACTA OTO-LARYNGOLOGICA, 1981, 91 (3-4) :223-235
[2]   DOSE MEASUREMENTS AND CALCULATIONS OF SMALL RADIATION-FIELDS FOR 9-MV X-RAYS [J].
ARCOVITO, G ;
PIERMATTEI, A ;
DABRAMO, G ;
BASSI, FA .
MEDICAL PHYSICS, 1985, 12 (06) :779-784
[3]   TREATMENT OF PINEALOMAS BY STEREOTAXIC RADIATION SURGERY [J].
BACKLUND, EO ;
RAHN, T ;
SARBY, B .
ACTA RADIOLOGICA-THERAPY PHYSICS BIOLOGY, 1974, 13 (04) :368-376
[4]  
BACKLUND EO, 1972, ACTA CHIR SCAND, V138, P749
[5]  
BARCIASALORIO JL, 1979, INSERM S, V12, P245
[6]   CRITICAL COMPARISON OF NEUROSURGICAL AND OTOLARYNGOLOGICAL APPROACHES TO ACOUSTIC NEUROMAS [J].
DITULLIO, MV ;
MALKASIAN, D ;
RAND, RW .
JOURNAL OF NEUROSURGERY, 1978, 48 (01) :1-12
[7]   SMALL-BEAM CALIBRATION BY 0.6-CM3 AND 0.2-CM3 IONIZATION CHAMBERS [J].
ENGLER, MJ ;
JONES, GL .
MEDICAL PHYSICS, 1984, 11 (06) :822-826
[8]   STEREOTACTIC HEAVY-ION BRAGG PEAK RADIOSURGERY FOR INTRA-CRANIAL VASCULAR DISORDERS - METHOD FOR TREATMENT OF DEEP ARTERIOVENOUS-MALFORMATIONS [J].
FABRIKANT, JI ;
LYMAN, JT ;
HOSOBUCHI, Y .
BRITISH JOURNAL OF RADIOLOGY, 1984, 57 (678) :479-490
[9]  
HIRCH A, 1979, ACTA OTOLARYNGOL, V88, P155
[10]  
HOUDEK PV, 1982, MED PHYS, V10, P333