INTRINSIC PEEP DETERMINED BY STATIC PRESSURE-VOLUME CURVES - APPLICATION OF A NOVEL AUTOMATED OCCLUSION METHOD

被引:5
作者
SYDOW, M
BURCHARDI, H
ZINSERLING, J
CROZIER, TA
DENECKE, T
ZIELMANN, S
机构
[1] Zentrum für Anaesthesiologie, Rettungs- und Intensivmedizin, Georg-August-Universität Göttingen, Göttingen, W-3400
关键词
INTRINSIC PEEP; EXTERNAL PEEP; STATIC COMPLIANCE; MECHANICAL VENTILATION; ALVEOLAR RECRUITMENT;
D O I
10.1007/BF01720534
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: Evaluation of new computer-controlled occlusion procedure for determination of intrinsic PEEP in mechanically ventilated patients and comparison with the standard end-expiratory occlusion method. Design: Prospective controlled study. Setting: Intensive care unit of a university hospital. Patients: 16 patients with acute respiratory failure of different degree and etiology. All patients were mechanically ventilated, heavily sedated and muscle paralyzed (non-depolarising relaxants). The type of ventilator, the inspiration/expiration ratio, FIO2 and PEEP were selected by the attending clinicians according to the patients' need and independently from the study. Interventions: Static compliance of the respiratory system (C(stat)) was determined at varying external end-expiratory pressure settings: ZEEP (= ambient pressure), PEEP of 5 cmH2O and 10 cmH2O. All other ventilator settings were kept constant during the entire procedure. Measurements and results: A computer-controlled occlusion method (SCASS) was used for determination of C(stat). Intrinsic PEEP was determined by SCASS as the extrapolated zero-volume intercept of the regression line of multiple pressure/volume data pairs (PEEP(SCASSinspir) and PEEP(SCASSexpir)). Directly thereafter intrinsic PEEP in this particular ventilatory setting was determined by end-expiratory occlusions (PEPP(EEO)). The intrinsic PEEP values of the different methods were nearly identical with a significant correlation (p<0.0001). Mean values+/-SD: PEEP(SCASSinspir) 7.1+/-4.3 cmH2O; PEEP(SCASSexpir) 7.1+/-4.5 cmH2O; PEEP(EEO) 7.1+/-4.2 cmH2O. Conclusion: Since no significant difference between PEEP(i) values measured by the inspiratory and expiratory occlusion method (SCASS) was seen, this indicates that no alveolar recruitment occurred during the respiratory cycle. This study demonstrates that the automated occlusion method for measuring C(stat) system can also be used with high accuracy for determination of intrinsic PEEP in mechanically ventilated patients.
引用
收藏
页码:166 / 171
页数:6
相关论文
共 21 条
[1]  
BAUM M, 1980, ANAESTHESIST, V29, P592
[2]  
BENITO S, 1985, INTENS CARE MED, V11, P76
[3]   RESPIRATORY COMPLIANCE AND RESISTANCE IN MECHANICALLY VENTILATED PATIENTS WITH ACUTE RESPIRATORY-FAILURE [J].
BERNASCONI, M ;
PLOYSONGSANG, Y ;
GOTTFRIED, SB ;
MILICEMILI, J ;
ROSSI, A .
INTENSIVE CARE MEDICINE, 1988, 14 (05) :547-553
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]  
BONE RC, 1983, RESPIR CARE, V28, P597
[6]   CAUSES OF ERROR OF RESPIRATORY PRESSURE-VOLUME CURVES IN PARALYZED SUBJECTS [J].
DALLAVASANTUCCI, J ;
ARMAGANIDIS, A ;
BRUNET, F ;
DHAINAUT, JF ;
CHELUCCI, GL ;
MONSALLIER, JF ;
LOCKHART, A .
JOURNAL OF APPLIED PHYSIOLOGY, 1988, 64 (01) :42-49
[7]   INTRINSIC PEEP ON STATIC PRESSURE-VOLUME CURVES [J].
FERNANDEZ, R ;
MANCEBO, J ;
BLANCH, L ;
BENITO, S ;
CALAF, N ;
NET, A .
INTENSIVE CARE MEDICINE, 1990, 16 (04) :233-236
[8]   THE ROLE OF TOTAL STATIC LUNG COMPLIANCE IN THE MANAGEMENT OF SEVERE ARDS UNRESPONSIVE TO CONVENTIONAL TREATMENT [J].
GATTINONI, L ;
PESENTI, A ;
CASPANI, ML ;
PELIZZOLA, A ;
MASCHERONI, D ;
MARCOLIN, R ;
IAPICHINO, G ;
LANGER, M ;
AGOSTONI, A ;
KOLOBOW, T ;
MELROSE, DG ;
DAMIA, G .
INTENSIVE CARE MEDICINE, 1984, 10 (03) :121-126
[9]   VOLUME PRESSURE CURVE OF TOTAL RESPIRATORY SYSTEM IN PARALYZED PATIENTS - ARTIFACTS AND CORRECTION FACTORS [J].
GATTINONI, L ;
MASCHERONI, D ;
BASILICO, E ;
FOTI, G ;
PESENTI, A ;
AVALLI, L .
INTENSIVE CARE MEDICINE, 1987, 13 (01) :19-25
[10]  
GOTTFRIED SB, 1985, AM REV RESPIR DIS, V131, P414