GLUCOSE BIOSENSOR BASED ON OXYGEN-ELECTRODE .1. SILASTIC COATED POLYCARBONATE MEMBRANES FOR BIOSENSOR APPLICATION

被引:9
作者
ATANASOV, P
WILKINS, E
机构
[1] Department of Chemical and Nuclear Engineering, University of New Mexico, Albuquerque, NM
关键词
GLUCOSE BIOSENSOR; OXYGEN ELECTRODE; SILASTIC COATING; POLYCARBONATE MEMBRANES;
D O I
10.1080/00032719308021482
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An amperometric glucose biosensor based on the oxygen electrode principle has been developed. Polycarbonate membranes (pore size from 0.01 mum to 0.4 mum) were used as external glucose diffusion membranes in order to obtain direct proportionality of the amperometric signal to the substrate concentration in the entire physiological range. The commercially available membranes - standard (hydrophilic, treated with Polyvinylpyrrolidone/(PVP)) and PVP-free membranes were compared with membranes coated with a silicone elastomer (silastic). Spin-drop coating technique was used to create stable, adhesive coatings over the polycarbonate membranes. These coated membranes achieved diffusion control of the glucose flux such that the amperometric signal of the biosensor was linearly proportional to the substrate concentration up to 16 mM glucose. The membrane parameters were optimized by varying the parameters of the coating process-spin rate of the membrane rotation and the silastic/water ratio in the coating emulsion.
引用
收藏
页码:1587 / 1612
页数:26
相关论文
共 30 条
[1]   DEVICES FOR THE CONTROL OF DIABETES-MELLITUS [J].
ALBISSER, AM .
PROCEEDINGS OF THE IEEE, 1979, 67 (09) :1308-1320
[2]   APPLICATION OF CHRONIC INTRAVASCULAR BLOOD-GLUCOSE SENSOR IN DOGS [J].
ARMOUR, JC ;
LUCISANO, JY ;
MCKEAN, BD ;
GOUGH, DA .
DIABETES, 1990, 39 (12) :1519-1526
[3]  
ATANAISOV P, 1991, CHEM COMM BULG ACAD, V24, P136
[4]   MASS-TRANSPORT AND REACTION KINETIC-PARAMETERS DETERMINED ELECTROCHEMICALLY FOR IMMOBILIZED GLUCOSE-OXIDASE [J].
CASTNER, JF ;
WINGARD, LB .
BIOCHEMISTRY, 1984, 23 (10) :2203-2210
[5]   ELECTRODE SYSTEMS FOR CONTINUOUS MONITORING IN CARDIOVASCULAR SURGERY [J].
CLARK, LC ;
LYONS, C .
ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1962, 102 (01) :29-&
[6]   IMPLANTED ELECTROENZYMATIC GLUCOSE SENSORS [J].
CLARK, LC ;
DUGGAN, CA .
DIABETES CARE, 1982, 5 (03) :174-180
[7]   THE DEVELOPMENT AND APPLICATION OF BIOSENSING DEVICES FOR BIOREACTOR MONITORING AND CONTROL [J].
CLARKE, DJ ;
CALDER, MR ;
CARR, RJG ;
BLAKECOLEMAN, BC ;
MOODY, SC ;
COLLINGE, TA .
BIOSENSORS, 1985, 1 (03) :213-320
[8]  
CLARKE WL, 1977, ARTIF ORGANS, V1, P78
[9]   ADVANCES IN BIOMEDICAL SENSOR TECHNOLOGY - A REVIEW OF THE 1985 PATENT LITERATURE [J].
DAVIS, G .
BIOSENSORS, 1986, 2 (02) :101-124
[10]   ULTRAMICRO OXYGEN ELECTRODE [J].
FATT, I .
JOURNAL OF APPLIED PHYSIOLOGY, 1964, 19 (02) :326-&