SURFACE ACOUSTIC-WAVE METHOD FOR INSITU DETERMINATION OF THE AMOUNTS OF ENZYME SUBSTRATE COMPLEX FORMED ON IMMOBILIZED GLUCOSE-OXIDASE DURING CATALYTIC REACTION

被引:21
作者
INOUE, Y [1 ]
KATO, Y [1 ]
SATO, K [1 ]
机构
[1] NAGAOKA UNIV TECHNOL,CTR ANAL,NAGAOKA,NIIGATA 94021,JAPAN
来源
JOURNAL OF THE CHEMICAL SOCIETY-FARADAY TRANSACTIONS | 1992年 / 88卷 / 03期
关键词
D O I
10.1039/ft9928800449
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Surface acoustic waves (SAWs) have been applied to the in situ determination of the amounts of an enzyme-substrate complex formed on an immobilized enzyme surface during the catalytic reaction. The input and output interdigital transducer (IDT) electrodes were photolithographically fabricated on a LiNbO3 crystal for the generation of Rayleigh SAWs and on LiTaO3 crystals for shear horizontal SAWs. The characteristics of propagation of these SAWs in contact with liquid phases were examined. Glucose oxidase (GOX) was immobilized between two IDT electrodes by silane coupling, the functionalization of which was characterized by X-ray photoelectron spectroscopy. The simultaneous determination of the reaction rates and of the adsorbed amounts of glucose was performed by combining an electrochemical method with the SAW method. When the concentration of glucose in the solution was changed, the amount of the GOX-glucose complex formed varied in rough proportion to the reaction rate, which gives experimental evidence for the establishment of pseudo-equilibrium between the adsorbed glucose and glucose in the solution and hence for the validity of the Michaelis-Menten equation. The reaction rate per enzyme-substrate complex was nearly constant, independent of the amount of glucose adsorbed. The formation of the enzyme-substrate complex showed a bell-shaped change with increase in pH, having a maximum at pH 7, which is quite similar to that observed for the dependence of reaction rate on pH. The catalytic functions of the immobilized active GOX were concluded to be almost equivalent.
引用
收藏
页码:449 / 454
页数:6
相关论文
共 13 条
[1]  
Calabrese G. S., 1986, IEEE 1986 Ultrasonics Symposium Proceedings (Cat. No.86CH2375-4), P607
[2]   SAW SENSORS [J].
DAMICO, A ;
VERONA, E .
SENSORS AND ACTUATORS, 1989, 17 (1-2) :55-66
[3]   A SURFACE ACOUSTIC-WAVE (SAW) PROBE FOR THE THERMOMECHANICAL CHARACTERIZATION OF SELECTED POLYMERS [J].
GROETSCH, JA ;
DESSY, RE .
JOURNAL OF APPLIED POLYMER SCIENCE, 1983, 28 (01) :161-178
[4]   CHANGES IN CATALYTIC ACTIVITIES OF CU THIN-FILMS BY SURFACE ACOUSTIC-WAVES PROPAGATED ON FERROELECTRIC LINBO3 [J].
INOUE, Y ;
MATSUKAWA, M .
JOURNAL OF THE CHEMICAL SOCIETY-CHEMICAL COMMUNICATIONS, 1990, (04) :296-297
[5]   EFFECT OF SURFACE ACOUSTIC-WAVE GENERATED ON FERROELECTRIC SUPPORT UPON CATALYSIS [J].
INOUE, Y ;
MATSUKAWA, M ;
SATO, K .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1989, 111 (24) :8965-8966
[6]  
Janata J., 1989, PRINCIPLES CHEM SENS, P69
[7]  
MARTIN SJ, 1990, SENSOR ACTUAT A-PHYS, V21, P712
[8]  
Moriizumi T., 1987, IEEE 1987 Ultrasonics Symposium Proceedings (Cat. No.87CH2492-7), P579
[9]  
Roberts R. F., 1980, Surface and Interface Analysis, V2, P5, DOI 10.1002/sia.740020103
[10]   MICROGRAVIMETRIC IMMUNOASSAY WITH PIEZOELECTRIC-CRYSTALS [J].