INTERCOMPARISON CAMPAIGN OF VERTICAL OZONE PROFILES INCLUDING ELECTROCHEMICAL SONDES OF ECC AND BREWER-MAST TYPE-A GROUND-BASED UV-DIFFERENTIAL ABSORPTION LIDAR

被引:57
作者
BEEKMANN, M [1 ]
ANCELLET, G [1 ]
MEGIE, G [1 ]
SMIT, HGJ [1 ]
KLEY, D [1 ]
机构
[1] RES CTR JULICH GMBH,INST CHEM POLLUTED ATMOSPHERE,W-5170 JULICH,GERMANY
关键词
VERTICAL OZONE PROFILES; TROPOSPHERIC OZONE RESEARCH PROJECT; BREWER-MAST SONDES; GROUND-BASED UV-DIAL;
D O I
10.1007/BF00694614
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
An intercomparison campaign was conducted at the Observatoire de Haute Provence (OHP) in Southern France in September 1989 in order to compare the three instruments used for vertical tropospheric ozone profiling in the European TOR (Tropospheric Ozone Research Project) network: balloon borne ECC and Brewer-Mast sondes and a ground based UV-DIAL (Differential Absorption Lidar). Additionally, a stratospheric lidar system and the Dobson spectrophotometer of the OHP were operated. Seven simultaneously measured vertical ozone profiles gave evidence for systematic differences of 15% between both types of electrochemical sondes in the troposphere, the Brewer-Mast sondes reading the smaller ozone values. These differences might be explained on the one hand by a possible contamination of the ozone sensor with reducing substances, causing a negative bias mainly for Brewer-Mast sondes and, on the other hand, by the evolution of the sonde background current during the flight, causing a positive bias for ECC sondes and a negative bias for Brewer-Mast sondes. The tropospheric lidar system, measuring the vertical ozone distribution between 6 and 12-15 km, showed ozone concentrations intermediate between the sonde results. This is in good agreement with its estimated systematic error of better than 7% in the upper troposphere. In the stratosphere, the differences between electrochemical sondes and the lidar are between 5 and 10% before the normalisation with the total ozone values measured by the Dobson spectrophotometer, and always below 5% after. While the Dobson normalisation thus corrects rather well the stratospheric part of the sonde profile, it only partially reduces errors occurring in the troposphere.
引用
收藏
页码:259 / 288
页数:30
相关论文
共 44 条
  • [1] OZONE PROFILE INTERCOMPARISON BASED ON SIMULTANEOUS OBSERVATIONS BETWEEN 20 AND 40 KM
    AIMEDIEU, P
    KRUEGER, AJ
    ROBBINS, DE
    SIMON, PC
    [J]. PLANETARY AND SPACE SCIENCE, 1983, 31 (07) : 801 - 807
  • [2] ANCELLET G, 1989, J ATMOS OPT TECH, V6, P832
  • [3] Attmannspacher W., 1981, BERICHTE DTSCH WETTE, V157
  • [4] Attmannspacher W., 1970, BERICHTE DTSCH WETTE, V120
  • [5] ELECTROCHEMICAL CONCENTRATION CELL OZONESONDE ACCURACY AND PRECISION
    BARNES, RA
    BANDY, AR
    TORRES, AL
    [J]. JOURNAL OF GEOPHYSICAL RESEARCH-ATMOSPHERES, 1985, 90 (ND5) : 7881 - 7887
  • [6] Bass A.M., 1985, P QUADRENNIAL OZONE, P606
  • [7] BECKMANN M, 1992, THESIS U PARIS 6
  • [8] THE OXFORD-KEW OZONE SONDE
    BREWER, AW
    MILFORD, JR
    [J]. PROCEEDINGS OF THE ROYAL SOCIETY OF LONDON SERIES A-MATHEMATICAL AND PHYSICAL SCIENCES, 1960, 256 (1287): : 470 - 495
  • [9] CHEDIN A, 1985, J CLIM APPL METEOROL, V24, P2
  • [10] CLAUDE H, 1987, WHO17 GLOB RES MON P