TRANSCRANIAL OPTICAL-PATH LENGTH IN INFANTS BY NEAR-INFRARED PHASE-SHIFT SPECTROSCOPY

被引:43
作者
BENARON, DA
KURTH, CD
STEVEN, JM
DELIVORIAPAPADOPOULOS, M
CHANCE, B
机构
[1] the Division of Neonatal and Developmental Medicine, Department of Pediatrics, Stanford University School of Medicine and the Hansen Free-Electron Laser Center at Stanford, and Lucile Salter Packard Hospital at Stanford, Stanford, CA
[2] From the Department of Anesthesiology, University of Pennsylvania School of Medicine and Children's Hospital of Philadelphia, Philadelphia, PA
[3] From the Department of Physiology, University of Pennsylvania School of Medicine and Children's Hospital of Philadelphia, Philadelphia, PA
[4] From the Department of Biochemistry/Biophysics, University of Pennsylvania School of Medicine and Children's Hospital of Philadelphia, Philadelphia, PA
来源
JOURNAL OF CLINICAL MONITORING | 1995年 / 11卷 / 02期
关键词
OXIMETRY; NEAR-INFRARED; SPECTROSCOPY; INFANT; PEDIATRIC; OPTICAL PATH LENGTH;
D O I
10.1007/BF01617732
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background. Near-infrared spectroscopy (NIRS) is an emerging technique for noninvasive, bedside monitoring of cerebral oxygenation and blood flow. Traditionally, it has relied on the Beer's Law relationship in which the concentration of light-absorbing oxygen-carrying pigments is proportional to their light absorbance, and inversely proportional to an optical path length (a measure of the distance traveled by photons passing through the tissue). In practice, NIRS has been based upon assumptions that mean transcranial optical path length, the average optical path length for a given patient, is constant among patients and independent of the wavelength of light used. Objective. The objective of our study was to measure mean optical transcranial path length in infants as a step in allowing quantitation of cerebral oxygenation. Methods. We measured mean transcranial optical path length in 34 infants, aged 1 day to 3 years, using amplitude-modulated phase-shift spectroscopy at 754 nm and 816 nm. Optical transcranial path lengths (mean +/- SEM) were 8.6 +/- 0.9 cm, 11.1 +/- 0.9 cm, and 11.3 +/- 0.9 cm at 754 nm, and 8.8 +/- 0.9 cm, 11.2 +/- 0.8 cm, and 11.1 +/- 0.9 cm at 816 nm, using emitter-detector separations of 1.8, 2.5, and 3.0 cm, respectively. Optical path length increased as emitter-detector separation, head circumference, or age increased. Variance in the ratio of mean optical path lengths at the two different wavelengths exceeded that accounted for by variation in repeated measures alone (p < 0.001), suggesting that optical path length is also not independent of wavelength. Conclusions. NIRS instrument emitter-detector geometry, subject age, head size, and wavelength used each influence optical path length. Quantitative NIRS measurements in clinical use may require concurrent measurement of both absorbance and optical path length at each wavelength, or use of newer measures that are not based upon Beer's Law assumptions.
引用
收藏
页码:109 / 117
页数:9
相关论文
共 49 条
[1]  
Jobsis F.F., Noninvasive infrared monitoring of cerebral and myocardial oxygen sufficiency and circulatory parameters, Science, 198, pp. 1264-1266, (1975)
[2]  
McCormick P.W., Stewart, Goetting M.G., Et al., Noninvasive cerebral optical spectroscopy for monitoring cerebral oxygen delivery and hemodynamics, Crit Care Med, 19, pp. 89-97, (1991)
[3]  
Brazy J.E., Effects of crying on cerebral blood volume and cytochrome aa<sub>3</sub>, J Pediatr, 112, 3, pp. 457-461, (1988)
[4]  
Edwards A.D., Richardson C., Cope M., Et al., Cotside measurement of cerebral blood flow in ill newborn infants by near-infrared spectroscopy, Lancet, 2, pp. 770-771, (1988)
[5]  
Wyatt J.S., Edwards A.D., Azzopardi D., Reynolds E.O.R., Magnetic resonance and near infrared spectroscopy for investigation of perinatal hypoxic-ischaemic brain injury, Arch Dis Child, 64, pp. 953-963, (1989)
[6]  
Wyatt J.S., Cope M., Delpy D.T., Et al., Quantitation of cerebral blood volume in human infants by near-infrared spectroscopy, J Appl Physiol, 68, 3, pp. 1086-1091, (1990)
[7]  
Benaron D.A., Kurth C.D., Steven J., Et al., Non-invasive estimation of cerebral oxygenation and oxygen consumption using phase-shift spectrophotometry, Proc IEEE Eng Med Biol, 12, 5, pp. 2004-2007, (1990)
[8]  
Edwards A.D., Wyatt J.S., Richardson C., Et al., Effects of indomethacin on cerebral haemodynamics in very preterm infants, Lancet, 335, pp. 1491-1495, (1990)
[9]  
Kurth C.D., Steven J.M., Benaron D., Chance B., Nearinfrared monitoring of the cerebral circulation, J Clin Monit, 9, 3, pp. 163-170, (1993)
[10]  
Benaron D.A., Stevenson D.K., Optical time-of-flight and absorbance imaging in biologic media, Science, 259, pp. 1463-1466, (1993)