REPEATABILITY OF A SPECIALLY DESIGNED INTRAORAL COLORIMETER

被引:60
作者
GOLDSTEIN, GR
SCHMITT, GW
机构
[1] New York University, College of Dentistry, New York, N. Y.
关键词
D O I
10.1016/0022-3913(93)90292-V
中图分类号
R78 [口腔科学];
学科分类号
1003 [口腔医学];
摘要
A specially modified Minolta CS-100 colorimeter with a DP-101 data processor was designed for intraoral use. To test the repeatability of the results obtained with this specially modified unit, five porcelain and five acrylic resin denture teeth were mounted in a fixed position in front of a fixed mounted meter. Three L*a*b* readings were taken per tooth and mean values were obtained after calibration as a baseline measurement. The readings were repeated on day 3 without recalibration and on day 8 after recalibration. The same-day repeatability showed DELTAE values <0.4 in nine of the 10 samples. The day 3 repeatability showed DELTAE values <0.4 in only 2 of the 10 samples. The day 8 results, after recalibration, showed DELTAE values >0.4 in all of the samples, with a range of 0.57 to 2.75. In vivo testing was performed on five patients with DELTAE values that ranged from 1.1 to 32.1.
引用
收藏
页码:616 / 619
页数:4