Error rate of multi-level rapid prototyping trajectories for pedicle screw placement in lumbar and sacral spine

被引:54
作者
Merc, Matjaz [1 ]
Drstvensek, Igor [2 ]
Vogrin, Matjaz [1 ]
Brajlih, Tomaz [2 ]
Friedrich, Tomaz [3 ]
Recnik, Gregor [1 ]
机构
[1] Univ Med Ctr Maribor, Dept Orthopaed Surg, Maribor, Slovenia
[2] Univ Maribor, Fac Mech Engn, Maribor, Slovenia
[3] Univ Med Ctr Maribor, Radiol Dept, Maribor, Slovenia
关键词
Drill guide; Template; Inaccuracy; Perforation; Radiation exposure;
D O I
10.3760/cma.j.issn.1008-1275.2014.05.003
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective: Free-hand pedicle screw placement has a high incidence of pedicle perforation which can be reduced with fluoroscopy, navigation or an alternative rapid prototyping drill guide template. In our study the error rate of multi-level templates for pedicle screw placement in lumbar and sacral regions was evaluated. Methods: A case series study was performed on 11 patients. Seventy-two screws were implanted using multilevel drill guide templates manufactured with selective laser sintering. According to the optimal screw direction preoperatively defined, an analysis of screw misplacement was performed. Displacement, deviation and screw length difference were measured. The learning curve was also estimated. Results: Twelve screws (17%) were placed more than 3.125 mm out of its optimal position in the centre of pedicle. The tip of the 16 screws (22%) was misplaced more than 6.25 mm out of the predicted optimal position. According to our predefined goal, 19 screws (26%) were implanted inaccurately. In 10 cases the screw length was selected incorrectly: 1 (1%) screw was too long and 9 (13%) were too short. No clinical signs of neurovascular lesion were observed. Learning curve was insignificantly noticeable (P=0.129). Conclusion: In our study, the procedure of manufacturing and applying multi-level drill guide templates has a 26% chance of screw misplacement. However, that rate does not coincide with pedicle perforation incidence and neurovascular injury. These facts along with a comparison to compatible studies make it possible to summarize that multi-level templates are satisfactorily accurate and allow precise screw placement with a clinically irrelevant mistake factor. Therefore templates could potentially represent a useful tool for routine pedicle screw placement.
引用
收藏
页码:261 / 266
页数:6
相关论文
共 28 条
[1]
Berlemann U, 1997, ORTHOPADE, V26, P463
[2]
Personalised image-based templates for intra-operative guidance [J].
Berry, E ;
Cuppone, M ;
Porada, S ;
Millner, PA ;
Rao, A ;
Chiverton, N ;
Seedhom, BB .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2005, 219 (H2) :111-118
[3]
Computer-assisted orthopedic surgery with individual templates and comparison to conventional operation method [J].
Birnbaum, K ;
Schkommodau, E ;
Decker, N ;
Prescher, A ;
Klapper, U ;
Radermacher, K .
SPINE, 2001, 26 (04) :365-369
[4]
Drstvensek I., 2008, INT J BIOL BIOMEDICA, V2, P29
[5]
ACCURACY OF PEDICLE SCREW PLACEMENT IN LUMBAR FUSIONS BY PLAIN RADIOGRAPHS AND COMPUTED-TOMOGRAPHY [J].
FARBER, GL ;
PLACE, HM ;
MAZUR, RA ;
JONES, DEC ;
DAMIANO, TR .
SPINE, 1995, 20 (13) :1494-1499
[6]
Accuracy of pedicle screw placement: a systematic review of prospective in vivo studies comparing free hand, fluoroscopy guidance and navigation techniques [J].
Gelalis, Ioannis D. ;
Paschos, Nikolaos K. ;
Pakos, Emilios E. ;
Politis, Angelos N. ;
Arnaoutoglou, Christina M. ;
Karageorgos, Athanasios C. ;
Ploumis, Avraam ;
Xenakis, Theodoros A. .
EUROPEAN SPINE JOURNAL, 2012, 21 (02) :247-255
[7]
Three-dimensional computed tomography-based, personalized drill guide for posterior cervical stabilization at C1-C2 [J].
Goffin, J ;
Van Brussel, K ;
Martens, K ;
Vander Sloten, J ;
Van Audekercke, R ;
Smet, MH .
SPINE, 2001, 26 (12) :1343-1347
[8]
Infection in the operating room [J].
Hughes, SPF ;
Anderson, FM .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1999, 81B (05) :754-755
[9]
Accuracy of computerized frameless stereotactic image-guided pedicle screw placement into previously fused lumbar spines [J].
Lim, MR ;
Girardi, FP ;
Yoon, SC ;
Huang, RC ;
Cammisa, FP .
SPINE, 2005, 30 (15) :1793-1798
[10]
Accuracy and efficacy of thoracic pedicle screws in scoliosis with patient-specific drill template [J].
Lu, Sheng ;
Zhang, Yuan Z. ;
Wang, Zheng ;
Shi, Ji H. ;
Chen, Yu B. ;
Xu, Xing M. ;
Xu, Yong Q. .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2012, 50 (07) :751-758