Several factors affecting the dissolution performance of various solid dosage forms tested using the now-through cell method have been evaluated in this study. These factors include the flow rate, the position of tablets in the flow through cell and the glass beads, as well as the physical properties of the dosage forms. The experimental results indicated that the flow rate through the cell greatly affects drug release from disintegrating tablets. Drug release increases with increasing the flow rate of the dissolution medium, as expected. However, the flow rate does not significantly influence drug release from tablets which are not disintegratable, for example, an erodible tablet. The position of the tablets in the flow cell without glass beads is also of importance. Drug release from horizontally positioned tablets is different from vertically positioned tablets in the flow-through cell. It was also observed that the use of glass beads in the flow cell can make the flow pattern more laminar. This may help avoid a turbulent agitation within the cell, which could impact on drug dissolution.
引用
收藏
页码:2063 / 2078
页数:16
相关论文
共 16 条
[1]   KINETICS OF INDOMETHACIN RELEASE FROM SUPPOSITORIES - INVITRO INVIVO CORRELATION [J].
AIACHE, JM ;
ISLASSE, M ;
BEYSSAC, E ;
AIACHE, S ;
RENOUX, R ;
KANTELIP, JP .
INTERNATIONAL JOURNAL OF PHARMACEUTICS, 1987, 39 (03) :235-242
[2]   APPARATUS FOR DETERMINING RATE OF DRUG RELEASE FROM SOLID DOSAGE FORMS [J].
BAUN, DC ;
WALKER, GC .
JOURNAL OF PHARMACEUTICAL SCIENCES, 1969, 58 (05) :611-&
[3]   DISSOLUTION STUDIES WITH A MULTICHANNEL CONTINUOUS-FLOW APPARATUS [J].
CAKIRYILDIZ, C ;
MEHTA, PJ ;
RAHMEN, W ;
SCHOENLEBER, D .
JOURNAL OF PHARMACEUTICAL SCIENCES, 1975, 64 (10) :1692-1697
[4]   EVALUATION OF A COLUMN-TYPE DISSOLUTION APPARATUS [J].
GROVES, MJ ;
ALKAN, MH ;
DEER, MA .
JOURNAL OF PHARMACY AND PHARMACOLOGY, 1975, 27 (06) :400-407
[5]  
KWEE HG, 1974, PHARM IND, V36, P576
[6]   IN VITRO ASSESSMENT OF DISSOLUTION KINETICS - DESCRIPTION AND EVALUATION OF A COLUMN-TYPE METHOD [J].
LANGENBUCHER, F .
JOURNAL OF PHARMACEUTICAL SCIENCES, 1969, 58 (10) :1265-+
[7]  
LANGENBUCHER F, 1989, PHARM IND, V51, P1276
[8]   DISSOLUTION RATE TESTING WITH COLUMN METHOD - METHODOLOGY AND RESULTS [J].
LANGENBUCHER, F ;
RETTIG, H .
DRUG DEVELOPMENT AND INDUSTRIAL PHARMACY, 1977, 3 (03) :241-263
[9]  
MARTIN A, 1983, PHYSICAL PHARM, pCH15
[10]   A COLLABORATIVE INVITRO DISSOLUTION STUDY USING THE FLOW-THROUGH METHOD [J].
NICKLASSON, M ;
WENNERGREN, B ;
LINDBERG, J ;
PERSSON, C ;
AHLGREN, R ;
PALM, B ;
PETTERSSON, A ;
WENNGREN, L .
INTERNATIONAL JOURNAL OF PHARMACEUTICS, 1987, 37 (03) :195-202