DEVELOPMENT OF A PATIENT-DEDICATED, ON-DEMAND, BLOOD-GAS MONITOR

被引:25
作者
MAHUTTE, CK
HOLODY, M
MAXWELL, TP
CHEN, PA
SASSE, SA
机构
[1] UNIV CALIF IRVINE,IRVINE,CA
[2] THREE M HEALTHCARE,CDI,TUSTIN,CA
关键词
D O I
10.1164/ajrccm.149.4.8143046
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
A new monitor (CDI(R) 2000) that brings blood gas measurements to the patient's bedside has been developed. To measure blood gases, blood is drawn into the patient's arterial pressure-monitoring line past in-line fluorescent-based sensors. After measurement, the blood is returned to the patient, avoiding blood loss and delays in sample turnaround and reducing the risk of infection to both patient and operator. We assessed this system's performance in vitro with tonometered bovine blood. Bias (mean difference between monitor and tonometered gas OF measured pH values) +/- the standard deviation (SD) were 0.01 +/- 0.02 at pH = 7.39; 0.0 +/- 0.7 mm Hg at PCO2 - 39 mm Hg; and 2.4 +/- 3.2 mm Hg at a PO2 = 100 mm Hg (n = 54). Changes in hematocrit, blood temperature, or serum sodium concentration did not have clinically significant effects on system performance. Studies in normal volunteers, in whom large changes in blood gases were induced, showed a bias (mean difference between monitor and IL 1306 values) +/- SD of 0.00 +/- 0.02 for pH, -0.4 +/- 2.0 mm Hg for PCO2, and -3.6 +/- 7.7 mm Hg for PO2 (n = 69). We conclude from the present study that the performance of this system is comparable to that of conventional blood gas analyzers.
引用
收藏
页码:852 / 859
页数:8
相关论文
共 29 条
[1]   MEASUREMENT IN MEDICINE - THE ANALYSIS OF METHOD COMPARISON STUDIES [J].
ALTMAN, DG ;
BLAND, JM .
JOURNAL OF THE ROYAL STATISTICAL SOCIETY SERIES D-THE STATISTICIAN, 1983, 32 (03) :307-317
[2]  
Altman DG, 1991, PRACTICAL STAT MED R, P206
[3]  
BUTTNER J, 1976, CLIN CHEM, V22, P532
[4]   OPTICAL FLUORESCENCE AND ITS APPLICATION TO AN INTRAVASCULAR BLOOD-GAS MONITORING-SYSTEM [J].
GEHRICH, JL ;
LUBBERS, DW ;
OPITZ, N ;
HANSMANN, DR ;
MILLER, WW ;
TUSA, JK ;
YAFUSO, M .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1986, 33 (02) :117-132
[5]  
HALBERT SA, 1990, CLIN CHEM, V36, P1581
[6]   COMPARISON OF BLOOD-GAS ANALYZER BIASES IN MEASURING TONOMETERED BLOOD AND A FLUOROCARBON-CONTAINING, PROFICIENCY-TESTING MATERIAL [J].
HANSEN, JE ;
JENSEN, RL ;
CASABURI, R ;
CRAPO, RO .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1989, 140 (02) :403-409
[7]  
HANSMANN DR, 1988, SPIE, V906, P4
[8]   IMPORTANCE OF CORRECT HANDLING OF SAMPLES FOR THE RESULTS OF BLOOD-GAS ANALYSIS [J].
HARSTEN, A ;
BERG, B ;
INEROT, S ;
MUTH, L .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1988, 32 (05) :365-368
[9]  
KESTEN S, 1991, CONT MANAGEMENT CRIT, V4, P119
[10]   EVALUATION OF A SINGLE TRANS-CUTANEOUS PO2-PCO2 SENSOR IN ADULT PATIENTS [J].
MAHUTTE, CK ;
MICHIELS, TM ;
HASSELL, KT ;
TRUEBLOOD, DM .
CRITICAL CARE MEDICINE, 1984, 12 (12) :1063-1066