POLYETHYLENE WEAR OF METAL-BACKED TIBIAL COMPONENTS IN TOTAL AND UNICOMPARTMENTAL KNEE PROSTHESES

被引:293
作者
ENGH, GA
DWYER, KA
HANES, CK
机构
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1992年 / 74卷 / 01期
关键词
D O I
10.1302/0301-620X.74B1.1732274
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We examined 86 polyethylene inserts, retrieved from total and unicompartmental knee prostheses after an average of 39.5 months in situ, grading them from 0 to 3 for seven modes of polyethylene degradation. Severe wear, with delamination or deformation, was observed in 51% of the implants, and was associated with time in situ, lack of congruency, thin polyethylene, third-body wear debris, and heat-pressed polyethylene. Significant under-surface cold flow was identified in some areas of unsupported polyethylene, and was associated with delamination in the load-bearing areas of thin inserts above screw holes in the underlying metal tray. We recommend the use of thicker polyethylene inserts, particularly in young, active patients and in designs with screw holes in the tibial baseplate. Thin polyethylene inserts which are at risk for accelerated wear and premature failure should be monitored radiographically at annual intervals.
引用
收藏
页码:9 / 17
页数:9
相关论文
共 20 条