RADIATION-DOSIMETRY BY AUTOMATIC IMAGE-ANALYSIS OF DICENTRIC CHROMOSOMES

被引:26
作者
BAYLEY, R
CAROTHERS, A
CHEN, X
FARROW, S
GORDON, J
JI, L
PIPER, J
RUTOVITZ, D
STARK, M
WALD, N
机构
[1] WESTERN GEN HOSP, MRC, HUMAN GENET UNIT, CREWE RD, EDINBURGH EH4 2XU, MIDLOTHIAN, SCOTLAND
[2] UNIV PITTSBURGH, GRAD SCH PUBL HLTH, DEPT ENVIRONM & OCCUPAT HLTH, PITTSBURGH, PA 15261 USA
来源
MUTATION RESEARCH | 1991年 / 253卷 / 03期
关键词
DICENTRIC CHROMOSOME; DOSE RESPONSE; GAMMA-DISTRIBUTION; IMAGE ANALYSIS; POISSON DISTRIBUTION; RADIATION DOSIMETRY;
D O I
10.1016/0165-1161(91)90135-U
中图分类号
Q3 [遗传学];
学科分类号
071007 ; 090102 ;
摘要
A system for scoring dicentric chromosomes by image analysis comprised fully automatic location of mitotic cells, automatic retrieval, focus and digitisation at high resolution, automatic rejection of nuclei and debris and detection and segmentation of chromosome clusters, automatic centromere location, and subsequent rapid interactive visual review of potential dicentric chromosomes to confirm positives and reject false positives. A calibration set of about 15000 cells was used to establish the quadratic dose response for Co-60 gamma-irradiation. The dose-response function parameters were established by a maximum likelihood technique, and confidence limits in the dose response and in the corresponding inverse curve, of estimated dose for observed dicentric frequency, were established by Monte Carlo techniques. The system was validated in a blind trial by analysing a test set comprising a total of about 8000 cells irradiated to 1 of 10 dose levels, and estimating the doses from the observed dicentric frequency. There was a close correspondence between the estimated and true doses. The overall sensitivity of the system in terms of the proportion of the total population of dicentrics present in the cells analysed that were detected by the system was measured to be about 40%. This implies that about 2.5 times more cells must be analysed by machine than by visual analysis. Taking this factor into account, the measured review time and false positive rates imply that analysis by the system of sufficient cells to provide the equivalent of a visual analysis of 500 cells would require about 1 h for operator review.
引用
收藏
页码:223 / 235
页数:13
相关论文
共 20 条
[1]  
Bauchinger M., 1984, BIOL DOSIMETRY, P15
[2]   CURRENT STATUS OF CYTOGENETIC PROCEDURES TO DETECT AND QUANTIFY PREVIOUS EXPOSURES TO RADIATION [J].
BENDER, MA ;
AWA, AA ;
BROOKS, AL ;
EVANS, HJ ;
GROER, PG ;
LITTLEFIELD, LG ;
PEREIRA, C ;
PRESTON, RJ ;
WACHHOLZ, BW .
MUTATION RESEARCH, 1988, 196 (02) :103-159
[3]   RAPID METAPHASE AND INTERPHASE DETECTION OF RADIATION-INDUCED CHROMOSOME-ABERRATIONS IN HUMAN-LYMPHOCYTES BY CHROMOSOMAL SUPPRESSION INSITU HYBRIDIZATION [J].
CREMER, T ;
POPP, S ;
EMMERICH, P ;
LICHTER, P ;
CREMER, C .
CYTOMETRY, 1990, 11 (01) :110-118
[4]   DISTRIBUTION OF THE VARIOUS RADIATION-INDUCED CHROMOSOMAL REARRANGEMENTS IN RELATION TO THE DOSE AND SAMPLING TIME [J].
DUTRILLAUX, B ;
VIEGASPEQUIGNOT, E ;
PRODHOMME, M ;
SPORTES, M .
MUTATION RESEARCH, 1985, 152 (2-3) :197-203
[5]  
GREEN DK, 1989, FLOW CYTOGENETICS, P161
[6]  
JENSEN RH, 1984, BIOL DOSIMETRY CYTOM, P161
[7]  
Ji L., 1989, Automation of cytogenetics., P177
[8]  
JI L, UNPUB AUTOMATIC CHRO
[9]   INTELLIGENT SPLITTING IN THE CHROMOSOME DOMAIN [J].
JI, LA .
PATTERN RECOGNITION, 1989, 22 (05) :519-532
[10]   MULTIPROCESSING INTERVAL PROCESSOR FOR AUTOMATED CYTOGENETICS [J].
LLOYD, D ;
PIPER, J ;
RUTOVITZ, D ;
SHIPPEY, G .
APPLIED OPTICS, 1987, 26 (16) :3356-3366