USE OF THE GRAVIMETRIC DILUTER IN MICROBIOLOGICAL WORK

被引:3
作者
MANNINEN, MT [1 ]
FUNG, DYC [1 ]
机构
[1] KANSAS STATE UNIV AGR & APPL SCI,DEPT ANIM SCI & IND,MANHATTAN,KS 66506
关键词
D O I
10.4315/0362-028X-55.1.59
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
Manual dilutions of food samples (e.g., 1:10 dilution) for microbiological investigations are cumbersome. We tested an automatic diluter (Model GD-600 Gravimetric Diluter, Spiral System, Inc., Bethesda, MD) for accuracy and sterility for microbiological work and found that the system provided +/- 0.1 to +0.2 g of accuracy in delivering a commanded amount of diluent. The delivered diluent was essentially sterile, with a few samples giving 1 to 4 colonies per ml after plating. With this system, weighing and dilution are fast, easy, and accurate.