PERCEPTUAL COMPARISON OF PULSED AND CONTINUOUS FLUOROSCOPY

被引:106
作者
AUFRICHTIG, R
XUE, P
THOMAS, CW
GILMORE, GC
WILSON, DL
机构
[1] CASE WESTERN RESERVE UNIV,DEPT PSYCHOL,CLEVELAND,OH 44106
[2] UNIV HOSP CLEVELAND,DEPT RADIOL,CLEVELAND,OH 44106
[3] CASE WESTERN RESERVE UNIV HOSP,DEPT RADIOL,CLEVELAND,OH 44106
关键词
MEDICAL IMAGING; X-RAY FLUOROSCOPY; VISUAL-PERCEPTION; X-RAY DOSE;
D O I
10.1118/1.597285
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Pulsed fluoroscopy (hereafter called pulsed) at reduced acquisition rates, typically 15 acq/s (pulsed-15), is proposed to reduce x-ray dose in interventional procedures. However, since the human visual system (HVS) acts as a temporal low-pass filter that interacts with such acquisitions, the proper dose for pulsed must be obtained in perception experiments. We determine the dose for low-frame-rate pulsed that gives visualization equivalent to that of conventional 30 acq/s fluoroscopy, hereafter called continuous. Computer-generated phantoms are used. They consist of stationary, low-contrast disks on a Aat background containing Poisson noise that mimics quantum noise in fluoroscopy. Image sequences are displayed on the video tachistoscope, a device with considerable display flexibility. Three experimental paradigms are used. (1) In a paired-comparison study, pulsed and continuous are displayed side-by-side on the same monitor, and the visibility of a contrast detail phantom is compared. (2) Using this same display, subjects record the minimally detectable disk contrast (the min-contrast measurement). (3) In a four-alternative forced-choice experiment, a disk is placed in one of four positions, and the subject determines the position of the disk. The methods are complementary-the forced-choice experiment properly eliminates the subjectivity of the observer threshold while the paired-comparison study is much more time efficient. With regard to pulsed and continuous comparisons, remarkable similarity is found between the supra-threshold experiments (1 and 2) and the detectability experiment (3); i.e., the average absolute differences in the equivalent-perception dose as determined by the three measures is approximately 3%. No difference is found between interlaced and noninterlaced display. A. relatively small dependence of dose savings on disk size is found with larger disks giving increased dose savings. Average dose savings of 22%, 38%, and 49% are found for pulsed-15, pulsed-10, and pulsed-7.5, respectively.
引用
收藏
页码:245 / 256
页数:12
相关论文
共 36 条
[1]  
[Anonymous], 1989, HUMAN MEMORY COGNITI
[2]  
[Anonymous], 1971, PROBIT ANAL
[3]  
[Anonymous], 1990, SPSS REFERENCE GUIDE
[4]  
AUFRICHTIG R, 1992, P SOC PHOTO-OPT INS, V1653, P374, DOI 10.1117/12.59515
[5]  
AUFRICHTIG R, 1992, RADIOLOGY, V185, P305
[6]  
AUFRICHTIG R, UNPUB MODEL PERCEPTI
[7]  
BERTHELSEN B, 1991, ACTA RADIOL, V32, P492
[8]   RADIATION EXPOSURE OF PERSONNEL DURING DIGITAL SUBTRACTION ANGIOGRAPHY [J].
BRITTON, CA ;
WHOLEY, MH .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 1988, 11 (02) :108-110
[9]   VISUAL SIGNAL-DETECTION .2. SIGNAL-LOCATION IDENTIFICATION [J].
BURGESS, AE ;
GHANDEHARIAN, H .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (08) :906-910
[10]   RADIATION EXPOSURE TO PATIENTS UNDERGOING PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY [J].
CASCADE, PN ;
PETERSON, LE ;
WAJSZCZUK, WJ ;
MANTEL, J .
AMERICAN JOURNAL OF CARDIOLOGY, 1987, 59 (09) :996-997