IN-PHANTOM CALIBRATION OF SELECTRON-LDR SOURCES

被引:7
作者
MEERTENS, H
机构
关键词
Brachytherapy; Quality assurance; accuracy; Reference air kerma rate; Source calibration;
D O I
10.1016/0167-8140(90)90010-T
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Source strength measurements were performed for cesium-137 spherical sources of nine Selectron-LDR remote afterloading systems in The Netherlands. The mean reference air kerma rate of a set of sources was obtained from air kerma rate measurements in a phantom at a distance of about 5.5 cm from a large number (24 to 30) of sources. The results were compared with the source strengths specified by the manufacturer. Discrepancies between measured and manufacturers data ranging from - 2.3 % to + 3.9 were observed. The frequency distribution of the strength of the sources within a set was measured by the use of a well-type ionisation chamber. The root mean square deviations of the nine source sets ranged from 1.0 to 3.2%. Application of the in-phantom calibration method would improve the precision of source strength measurements and therefore reduce the differences in dose delivery between institutes. The uncertainty of the in-phantom method, which could be estimated by statistical methods, was 0.3 (one standard deviation). That part of the uncertainty, which could not be evaluated by statistical methods, was estimated to be about 1.2% (one effective standard deviation). This latter part in the uncertainty needs further investigation in order to reduce the overall uncertainty. © 1990.
引用
收藏
页码:369 / 378
页数:10
相关论文
共 12 条
[2]   PHOTON MASS ATTENUATION AND ENERGY-ABSORPTION COEFFICIENTS FROM 1 KEV TO 20 MEV [J].
HUBBELL, JH .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1982, 33 (11) :1269-1290
[3]  
JONES CH, 1988, DOSIMETRY RADIOTHERA, V1, P275
[4]  
KAARLS R, 1981, P VERB COM INT POIDS, P49
[5]   BRACHYTHERAPY BUILDUP FACTORS [J].
KORNELSEN, RO ;
YOUNG, MEJ .
BRITISH JOURNAL OF RADIOLOGY, 1981, 54 (638) :136-136
[6]  
LOEVINGER R, 1981, MED PHYSICS MONOGRAP, V7, P22
[7]   EFFECTIVE ATTENUATION IN WATER OF GAMMA RAYS OF GOLD 198 IRIDIUM 192 CESIUM 137 RADIUM 226 AND COBALT 60 [J].
MEISBERGER, LL ;
KELLER, RJ ;
SHALEK, RJ .
RADIOLOGY, 1968, 90 (05) :953-+
[8]   CONSISTENCY AND SIMPLICITY IN THE DETERMINATION OF ABSORBED DOSE TO WATER IN HIGH-ENERGY PHOTON BEAMS - A NEW CODE OF PRACTICE [J].
MIJNHEER, BJ ;
AALBERS, AHL ;
VISSER, AG ;
WITTKAMPER, FW .
RADIOTHERAPY AND ONCOLOGY, 1986, 7 (04) :371-384
[9]   DOSE IN WATER SURROUNDING POINT ISOTROPIC GAMMA-RAY EMITTERS [J].
WEBB, S ;
FOX, RA .
BRITISH JOURNAL OF RADIOLOGY, 1979, 52 (618) :482-484
[10]   THE PRECISION OF DOSE CALCULATION AND THE ACTIVITY DISTRIBUTION OF SELECTRON SOURCES [J].
WILKINSON, JM .
BRITISH JOURNAL OF RADIOLOGY, 1982, 55 (658) :778-779