INVESTIGATION OF CHEMICAL MODIFICATION IN THE SIMULTANEOUS GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRIC DETECTION OF ARSENIC, CADMIUM, LEAD, AND SELENIUM
SIMULTANEOUS GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROMETRY;
CHEMICAL MODIFICATION;
D O I:
10.1080/00387019408000841
中图分类号:
O433 [光谱学];
学科分类号:
0703 ;
070302 ;
摘要:
The use of a combined palladium nitrate-magnesium nitrate chemical modifier allows a compromise higher pyrolysis temperature to be used for the simultaneous detection of arsenic, cadmium, lead, and selenium by graphite furnace atomic absorption spectrometry. For arsenic and selenium 900-degrees-C, for arsenic, selenium, and lead 600-degrees-C, and for arsenic, cadmium, lead, and selenium 350-degrees-C are recommended.