BIOPSY NEEDLES AND DEVICES - ASSESSMENT OF FERROMAGNETISM AND ARTIFACTS DURING EXPOSURE TO A 1.5-T MR SYSTEM

被引:25
作者
MOSCATEL, MA [1 ]
SHELLOCK, FG [1 ]
MORISOLI, SM [1 ]
机构
[1] CHICAGO MED SCH,CHICAGO,IL
来源
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING | 1995年 / 5卷 / 03期
关键词
ARTIFACT; BIOPSIES; TECHNOLOGY; MAGNETIC RESONANCE (MR) GUIDANCE; SAFETY;
D O I
10.1002/jmri.1880050324
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Forty-eight biopsy devices (needles, stylets, guide wires, biopsy guns, etc) were evaluated for ferromagnetism (n = 48) and artifacts (n = 43) in a 1.5-T magnetic resonance (MR) system, with previously described techniques. Forty-five of the 48 devices were found to be ferromagnetic. Artifacts varied, depending on the type and size of the component material(s) of the device. Most of the commercially available biopsy devices are not useful for MR imaging-guided biopsy procedures because of the presence of ferromagnetism and associated artifacts.
引用
收藏
页码:369 / 372
页数:4
相关论文
共 19 条
  • [1] Shellock FG, Kanal E, Magnetic resonance: bioeffects, safety and patient management, (1994)
  • [2] Shellock FG, Schatz CJ, Metallic otologic implants: in vitro assessment of ferromagnetism at 1.5 T, AJNR, 12, pp. 279-281, (1991)
  • [3] Shellock FG, Crues JV, High‐field strength MR imaging and metallic biomedical implants: an ex vivo evaluation of deflection forces, AJR, 151, pp. 389-392, (1988)
  • [4] Shellock FG, Morlsoli SM, Kanal E, MR procedures and blomedical implants, materials, and devices: 1993 update, Radiology, 189, pp. 587-599, (1993)
  • [5] Yuh WTC, Hanlgan MT, Nerad JA, Et al., Extrusion of eye socket magnetic Implant after MR imaging: potential hazard to patient with eye prosthesis, JMRI, 1, pp. 711-713, (1991)
  • [6] Shellock FG, Morlsoli SM, Ex vivo evaluation of ferromagnetism and artifacts of cardiac occluders exposed to a 1.5‐T MR system, JMRI, 4, pp. 213-215, (1994)
  • [7] Lufkin R, Layfield L, Coaxial needle system of MR‐ and CT‐guided aspiration cytology, J Comput Assist Tomogr, 13, pp. 1105-1107, (1989)
  • [8] Lufkin R, Teresi L, Hanafee W, New needle for MR‐guided aspiration cytology of the head and neck, AJR, 149, pp. 380-382, (1987)
  • [9] Duckwlier G, Lufkin RB, Teresi L, Et al., Head and neck lesions: MR‐guided aspiration biopsy, Radiology, 170, pp. 519-522, (1989)
  • [10] Lufkin R, Teresi L, Chiu L, Hanafee W, A technique for MR‐guided needle placement, AJR, 151, pp. 193-196, (1988)