INHIBITION OF A DEMAND PACEMAKER AND INTERFERENCE WITH MONITORING EQUIPMENT BY RADIO-FREQUENCY TRANSMISSIONS

被引:22
作者
PICKERS, BA
GOLDBERG, MJ
机构
[1] Leicester Regional Cardiothoracic Centre, Groby Road Hospital, Leicester
关键词
D O I
10.1136/bmj.2.5655.504
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
During the initial testing of Radio Leicester a swept-frequency technique for testing radio antennae was shown to affect demand pacemakers by inhibition of the pacing impulse and to interfere with physiological monitoring equipment. Adequate filtering of demand pacemakers is necessary to eliminate this interference. There is no evidence that such testing has any effect on the function of fixed-rate pacemakers. There is also a potential danger to implanted demand pacemakers by the use of similar polyscope generators used for servicing radio and television apparatus in industry. © 1969, British Medical Journal Publishing Group. All rights reserved.