VERSATILE TYPE MINIATURE DIAMOND ANVIL HIGH-PRESSURE CELL

被引:26
作者
YAMAOKA, S
FUKUNAGA, O
SHIMOMURA, O
NAKAZAWA, H
机构
[1] National Institute for Researches in Inorganic Materials, Niihari-gun
关键词
D O I
10.1063/1.1136017
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A miniature diamond anvil high-pressure cell 42 mm in diameter, 18 mm thick, and 150 g in weight has been constructed. The cell is loaded into a piston-cylinder type pressing device with an alignment mechanism and compressed to a desired pressure. After the pressure is clamped by three screws, the cell part is taken out. Thereby, in spite of the small and simple cell, pressure can be generated easily and steadily up to 200 kbar. Because the cell is small and light enough to be mounted on a usual goniometer head, and has large openings towards both the top and bottom, it can widely be used as a versatile apparatus for research at high pressure.
引用
收藏
页码:1163 / 1164
页数:2
相关论文
共 3 条
[1]   XRAY DIFFRACTION AND OPTICAL OBSERVATIONS ON CRYSTALLINE SOLIDS UP TO 300 KBAR [J].
BASSETT, WA ;
TAKAHASHI, T ;
STOOK, PW .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1967, 38 (01) :37-+
[2]   CALIBRATION OF PRESSURE-DEPENDENCE OF R1 RUBY FLUORESCENCE LINE TO 195 KBAR [J].
PIERMARINI, GJ ;
BLOCK, S ;
BARNETT, JD ;
FORMAN, RA .
JOURNAL OF APPLIED PHYSICS, 1975, 46 (06) :2774-2780
[3]  
TAKEMURA K, UNPUBLISHED