A 4-CHANNEL TIME-DOMAIN MULTIPLEXER - A COST-EFFECTIVE ALTERNATIVE TO MULTIPLE RECEIVERS

被引:37
作者
PORTER, JR
WRIGHT, SM
FAMILI, N
机构
[1] Department of Electrical Engineering, Texas A&M University, College Station, Texas
关键词
RF MULTIPLEXING; MR ARRAY COILS; BAND-PASS FILTERING; MULTICHANNEL ACQUISITION;
D O I
10.1002/mrm.1910320412
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The concept of simultaneous image acquisition as originally conceived by Hyde has been implemented by Roemer et al. using a multiple receiver system. This article describes an alternative technique that uses time domain multiplexing (TDM) to acquire simultaneous images using a single receiver channel. This method requires few modifications to the existing receiver and has been shown to be equivalent to a true four channel receiver in most applications. The multiplexing technique was implemented and tested on a standard commercial scanner that has also been equipped with a four channel receiver. Signal-to noise results demonstrate that four independent images can be multiplexed through a single receiver channel with no degradation in image quality. Additionally, spine images that were obtained from a normal volunteer with both a multiplexed single channel receiver and a true four channel receiver system are presented.
引用
收藏
页码:499 / 504
页数:6
相关论文
共 12 条
[1]  
FRANKS LE, 1981, SIGNAL THEORY, P73
[2]   NOISE CORRELATIONS IN DATA SIMULTANEOUSLY ACQUIRED FROM MULTIPLE SURFACE COIL ARRAYS [J].
HAYES, CE ;
ROEMER, PB .
MAGNETIC RESONANCE IN MEDICINE, 1990, 16 (02) :181-191
[3]   PARALLEL IMAGE ACQUISITION FROM NONINTERACTING LOCAL COILS [J].
HYDE, JS ;
JESMANOWICZ, A ;
FRONCISZ, W ;
KNEELAND, JB ;
GRIST, TM ;
CAMPAGNA, NF .
JOURNAL OF MAGNETIC RESONANCE, 1986, 70 (03) :512-517
[4]  
MUMFORD WW, 1968, NOISE PERFORMANCE FA, pCH6
[5]  
PORTE JR, 1993, THESIS TEXAS A M U
[6]  
PORTER JR, 1992, P SMRI ANN M NEW YOR, P154
[7]  
PORTER JR, 1992, 11TH P SMRM ANN M NE, P4007
[8]  
PORTER JR, 1992, 11TH P SMRM ANN M NE, P4012
[9]   THE NMR PHASED-ARRAY [J].
ROEMER, PB ;
EDELSTEIN, WA ;
HAYES, CE ;
SOUZA, SP ;
MUELLER, OM .
MAGNETIC RESONANCE IN MEDICINE, 1990, 16 (02) :192-225
[10]  
WRIGHT SM, 1993, 12TH P SMRM ANN M NE, P933