THE EFFECT OF SECONDARY-ELECTRON SPREAD ON THE PENUMBRA IN HIGH-ENERGY PHOTON-BEAM THERAPY

被引:15
作者
DAY, MJ
LAMBERT, GD
LOCKS, SM
机构
关键词
D O I
10.1259/0007-1285-63-748-278
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In photon beam therapy, the geometric penumbra width is determined by the source-size and the collimator design. The width of the physical (i.e. dosimetric) penumbra involves an additional contribution from secondary electron spread. Using a suitably defined measure of penumbra width, the separate widths due to photons and secondary electrons are additive in quadrature. Secondary electron spread functions were measured using photographic dosimetry for 60Co γ rays and for 4, 8, and 16 MV X rays. The results suggest that, under typical treatment conditions, secondary electron spread may be the predominant contributor to the penumbra at effective generating voltages exceeding 10 MV. The implications for selection of beam energy in precision small-field radiotherapy are discussed briefly.
引用
收藏
页码:278 / 285
页数:8
相关论文
共 34 条
[1]  
ADAMS DG, 1964, RADIOLOGY, V83, P785
[2]   A METHOD FOR CALCULATING MEGAVOLTAGE X-RAY DOSE AND DOSE PARAMETERS [J].
BAGNE, F .
MEDICAL PHYSICS, 1980, 7 (06) :664-672
[3]  
BJARNGARD BE, 1986, MED PHYS, V13, P760
[4]   CALCULATION OF PHOTON DOSE DISTRIBUTIONS IN AN INHOMOGENEOUS-MEDIUM USING CONVOLUTIONS [J].
BOYER, AL ;
MOK, EC .
MEDICAL PHYSICS, 1986, 13 (04) :503-509
[5]   EXTRACTION OF PENCIL BEAM KERNELS BY THE DECONVOLUTION METHOD [J].
CHUI, CS ;
MOHAN, R .
MEDICAL PHYSICS, 1988, 15 (02) :138-144
[6]  
CUNNINGHAM J R, 1972, Computer Programs in Biomedicine, V2, P192, DOI 10.1016/0010-468X(72)90029-3
[7]   SCATTER-AIR RATIOS [J].
CUNNINGHAM, JR .
PHYSICS IN MEDICINE AND BIOLOGY, 1972, 17 (01) :42-+
[8]   ANALYSIS OF PHYSICAL PARAMETERS ASSOCIATED WITH THE MEASUREMENT OF HIGH-ENERGY X-RAY PENUMBRA [J].
DAWSON, DJ ;
HARPER, JM ;
AKINRADEWO, AC .
MEDICAL PHYSICS, 1984, 11 (04) :491-497
[9]   SPECIFICATION AND ADDITIVITY OF UNSHARPNESS IN DIAGNOSTIC RADIOLOGY [J].
DAY, MJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (03) :399-405
[10]   THE 4-MEV LINEAR ACCELERATOR AT NEWCASTLE UPON TYNE [J].
DAY, MJ ;
FARMER, FT .