REAL-TIME MICROCONTROLLER IMPLEMENTATION OF AN ADAPTIVE MYOELECTRIC FILTER

被引:4
作者
BAGWELL, PJ [1 ]
CHAPPELL, PH [1 ]
机构
[1] UNIV SOUTHAMPTON,DEPT ELECT ENGN,SOUTHAMPTON SO17 1BJ,HANTS,ENGLAND
关键词
ADAPTIVE FILTER; EMG; MICROCONTROLLER; MUSCLE FORCE; MYOELECTRIC CONTROL;
D O I
10.1016/1350-4533(95)91888-N
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper describes a real time digital adaptive filter for processing myoelectric signals. The filter time constant is automatically selected by the adaption algorithm, giving a significant improvement over linear filters for estimating the muscle force and controlling a prosthetic device. Interference from mains sources often produces problems for myoelectric processing, and so 50 Hz and all harmonic frequencies are reduced by an averaging filter and differential process. This makes practical electrode placement and contact less critical and lime consuming An economic real time implementation is essential for a prosthetic controller, and this is achieved using an Intel 80C196KC microcontroller.
引用
收藏
页码:151 / 160
页数:10
相关论文
共 13 条
[1]  
AKKIRAJU P, IEEE T BIOMED ENG, V39, P652
[2]   PHYSIOLOGY AND MATHEMATICS OF MYOELECTRIC SIGNALS [J].
DE LUCA, CJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1979, 26 (06) :313-325
[3]   SIGNAL-PROCESSING FOR PROPORTIONAL MYOELECTRIC CONTROL [J].
EVANS, HB ;
PAN, Z ;
PARKER, PA ;
SCOTT, RN .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1984, 31 (02) :207-211
[4]   A LEARNING FILTER FOR REMOVING NOISE INTERFERENCE [J].
FURNO, GS ;
TOMPKINS, WJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1983, 30 (04) :234-235
[6]   MYOELECTRIC SIGNAL-PROCESSING - OPTIMAL ESTIMATION APPLIED TO ELECTROMYOGRAPHY .1. DERIVATION OF THE OPTIMAL MYOPROCESSOR [J].
HOGAN, N ;
MANN, RW .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1980, 27 (07) :382-395
[7]  
KYBERD PJ, 1992, JUN INT SOC PROSTH O
[8]   COMPARISON OF SIGNAL-TO-NOISE RATIO OF MYOELECTRIC FILTERS FOR PROSTHESIS CONTROL [J].
MEEK, SG ;
FETHERSTON, SJ .
JOURNAL OF REHABILITATION RESEARCH AND DEVELOPMENT, 1992, 29 (04) :9-20
[9]   A NOTE ON THE TIME CONSTANT IN LOW-PASS FILTERING OF RECTIFIED SURFACE EMG [J].
MIYANO, H ;
MASUDA, T ;
SADOYAMA, T .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1980, 27 (05) :274-278
[10]   CLINICAL-APPLICATION OF AN ACTIVE ELECTRODE USING AN OPERATIONAL-AMPLIFIER [J].
NISHIMURA, S ;
TOMITA, Y ;
HORIUCHI, T .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1992, 39 (10) :1096-1099