DISTORTIONS FROM CURVED FLOW IN MAGNETIC-RESONANCE-IMAGING

被引:13
作者
FRANK, LR
BUXTON, RB
机构
[1] Magnetic Resonance Institute, Department of Radiology, University of California at San Diego, San Diego, California
关键词
D O I
10.1002/mrm.1910290115
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We present an analysis of how vessel curvature can create distortions in magnetic resonance images of flowing blood. Steady flow in curved vessels produces distortions of the vessel shape and intensity variations in the image due to motion during the interval between phase encoding or slice selection and the echo center. Even with steady flow, vessel curvature produces motion moments higher than velocity (acceleration, etc.), but use of a first order oblique flow compensated phase encoding gradient waveform reduced the distortion in the image. Numerical calculations of image distortions based on simple flow models are in good agreement with experimental results in a phantom.
引用
收藏
页码:84 / 93
页数:10
相关论文
共 18 条
  • [1] THE MOTION ARTIFACT SUPPRESSION TECHNIQUE (MAST) IN MAGNETIC-RESONANCE IMAGING - CLINICAL-RESULTS
    COLLETTI, PM
    RAVAL, JK
    BENSON, RC
    PATTANY, PM
    ZEE, CS
    BOSWELL, WD
    NORRIS, SL
    RALLS, PW
    SEGALL, HD
    [J]. MAGNETIC RESONANCE IMAGING, 1988, 6 (03) : 293 - 299
  • [2] EDELMAN RR, 1990, CLIN MAGNETIC RESONA
  • [3] ELIMINATION OF OBLIQUE FLOW ARTIFACTS IN MAGNETIC-RESONANCE-IMAGING
    FRANK, LR
    CRAWLEY, AP
    BUXTON, RB
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1992, 25 (02) : 299 - 307
  • [4] FRANK LR, 1991, 9TH SMRI ANN M, P323
  • [5] HAACKE E M, 1986, Magnetic Resonance Imaging, V4, P359, DOI 10.1016/0730-725X(86)91046-5
  • [6] SPATIAL MISREGISTRATION OF VASCULAR FLOW DURING MR-IMAGING OF THE CNS - CAUSE AND CLINICAL-SIGNIFICANCE
    LARSON, TC
    KELLY, WM
    EHMAN, RL
    WEHRLI, FW
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1990, 155 (05) : 1117 - 1124
  • [7] MR ANGIOGRAPHY WITH GRADIENT MOTION REFOCUSING
    LAUB, GA
    KAISER, WA
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1988, 12 (03) : 377 - 382
  • [8] INPLANE VASCULAR IMAGING - PULSE SEQUENCE DESIGN AND STRATEGY
    LENZ, GW
    HAACKE, EM
    MASARYK, TJ
    LAUB, G
    [J]. RADIOLOGY, 1988, 166 (03) : 875 - 882
  • [9] MORAN P R, 1987, Magnetic Resonance Imaging, V5, P34
  • [10] MAGNETIC-RESONANCE ANGIOGRAPHY
    NISHIMURA, DG
    MACOVSKI, A
    PAULY, JM
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1986, 5 (03) : 140 - 151