LIF-MG, CU, P TL DETECTORS WITH ADJUSTABLE ENERGY RESPONSES

被引:2
作者
WANG, ST
CAI, GG
机构
[1] Solid Dosimetric Detector and Method Laboratory, Beijing, 102205
关键词
D O I
10.1088/0022-3727/24/11/025
中图分类号
O59 [应用物理学];
学科分类号
摘要
When used for tissue-equivalent monitoring, in accordance with ICRU Reports 39 (1985) and 43 (1988), the thermoluminescence material LiF:Mg, Cu, P shows an 'under-response' to low-energy photons. This paper explores a new way to increase its low-energy response effectively, and to make it adjustable to some extent, by adding elements with high Z(eff). At the same time its excellent TL performance is maintained. Different concentrations of heavy elements are needed to satisfy the requirements for the ICRU quantities H(p)(10) and H*(10).
引用
收藏
页码:2072 / 2075
页数:4
相关论文
共 9 条
[1]   LOW-TEMPERATURE AND HIGH-TEMPERATURE RESPONSE OF LITHIUM-FLUORIDE DOSIMETERS TO X-RAYS [J].
BUDD, T ;
MARSHALL, M ;
PEAPLE, LHJ ;
DOUGLAS, JA .
PHYSICS IN MEDICINE AND BIOLOGY, 1979, 24 (01) :71-80
[2]  
Horowitz Y, 1990, RADIAT PROT DOSIM, V30, P75
[3]   DEVELOPMENT OF A NEW HIGHLY SENSITIVE LIF THERMOLUMINESCENCE DOSIMETER AND ITS APPLICATIONS [J].
NAKAJIMA, T ;
MURAYAMA, Y ;
MATSUZAWA, T ;
KOYANO, A .
NUCLEAR INSTRUMENTS & METHODS, 1978, 157 (01) :155-162
[4]   THERMOLUMINESCENCE SYSTEM FOR INTERCOMPARISON OF ABSORBED DOSE AND RADIATION QUALITY OF X-RAYS WITH A HVL OF 0.1 TO 3.0MM CU [J].
PUITE, KJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (02) :216-225
[5]   USE OF PRECISION THERMOLUMINESCENCE DOSIMETRY FOR INTERCOMPARISON OF ABSORBED DOSE [J].
ROSSITER, MJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1975, 20 (05) :735-746
[6]  
Wagner S.R., 1985, RADIAT PROT DOSIM, V12, P231
[7]  
WANG SS, 1986, RADIAT PROT DOSIM, V14, P223
[8]  
1985, ICRU39 ICRU PUBL
[9]  
1988, ICRU43 ICRU PUBL