CONTINUOUS-FLOW SENSOR STRATEGY COMPRISING A ROTATING BIOREACTOR AND A STATIONARY RING AMPEROMETRIC DETECTOR

被引:25
作者
MATSUMOTO, K
BAEZA, JJB
MOTTOLA, HA
机构
[1] OKLAHOMA STATE UNIV,DEPT CHEM,STILLWATER,OK 74078
[2] KYUSHU UNIV,FAC AGR,DEPT FOOD SCI & TECHNOL,FUKUOKA 812,JAPAN
[3] UNIV VALENCIA,FAC MED,DEPT ANALYT CHEM,VALENCIA,SPAIN
关键词
D O I
10.1021/ac00053a025
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A rotating immobilized-enzyme reactor and a static Pt-ring electrode have been designed In a sealed cell for use under continuous-flow/stopped-flow operation. The operating characteristics of the integrated bioreactor/detector unit have been studied for determination of glucose using immobilized glucose oxidase [EC 1. 1.3.41 and amperometric monitoring of the H2O2 produced. Determinations are based on the rate of response under stopped-flow conditions. The trend in values of the apparent Michaelis-Menten constant confirms the validity of the approach to sensitive determinations utilizing very small amounts of biocatalysts.