CORNEAL MARKER FOR RUNNING SUTURES IN PENETRATING KERATOPLASTY

被引:5
作者
DUNCKER, GIW
NOLLE, B
机构
[1] Klinik fur Ophthalmologie Kiel, D-24105 Kiel
关键词
CORNEAL MARKER; DOUBLE RUNNING SUTURE; PENETRATING KERATOPLASTY;
D O I
10.1055/s-2008-1035349
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background In penetrating keratoplasty adaptation of wound edges, mode of trephination of both donor and patient cornea and last but not least techniques of suture style are the factors determining postoperative astigmatism. A new corneal marking device for keratoplasties and epikeratophakias is introduced. This marker helps the surgeon both to center the trephine and to perform a perfect double running torque antitorque suture. Methods The marker has the shape of an 8-bite star. It can be stained and printed on the cornea. The cross hair of the marker has to be right in the optical center. After trephination we fix the graft with temporal single sutures. Then we mark the dried cornea, so that an 8-bite star can be printed on the cornea. Now a first 10-0 running suture is done. The second continuous suture can be set just in between. In our experience it turns out more precisely, if this second running suture is previously marked, too. Results and conclusions After 35 perforating keratoplasties performed with this new marking device the mean astigmatism was 2.0+/-0.75 dpt (x+/-sx). Only a prospective, randomized study comparing different suture styles with and without star marker can elucidate the definite value of this new surgical instrument.
引用
收藏
页码:51 / 52
页数:2
相关论文
empty
未找到相关数据