Background. It has been suggested that left ventricular unloading at the time of reperfusion provides superior infarct salvage over reperfusion alone. The purpose of this study was to show that the Hemopump transvalvular axial-flow left ventricular assist device provides superior left ventricular unloading, ischemic zone collateral blood flow, and infarct size reduction compared with intra-aortic balloon counterpulsation and reperfusion alone. Methods and Results. Eighteen dogs were instrumented with regional myocardial function sonomicrometers in the ischemic and control zones. The left anterior descending coronary artery just distal to the first diagonal branch was instrumented with a silk snare and Doppler flow probe. Additionally, pressure catheters were placed in the left atrial appendage, left ventricular apex, and ascending aorta for hemodynamic measurements. Regional myocardial blood flow was determined by using 15-mu-m radioactive microspheres. Measurements were made in the control state, immediately after coronary occlusion, at 1 and 2 hours after coronary occlusion, with reperfusion, and 1 hour after reperfusion. In treated animals, left ventricular assistance was maintained during the entire period of occlusion and reperfusion. The Hemopump was associated with a significant decrease in left ventricular systolic and diastolic pressure, whereas mean arterial pressure was maintained. Intra-aortic balloon counterpulsation resulted in no significant changes in left ventricular systolic pressure and a modest decrease in left ventricular diastolic pressure. Regional unloading as assessed by sonomicrometers was significant in the Hemopump animals and absent in the balloon pump animals. Absolute regional myocardial blood flow in the ischemic zone increased slightly (p = 0.002) in the Hemopump animals and did not change in the balloon pump animals. Infarct size expressed as percentage of the zone at risk was 62.6% in the control animals, 27.22% in the balloon pump animals, and 21.7% in the Hemopump animals. Conclusions. Mechanical unloading of the ventricle during ischemia and reperfusion appears to result in significant infarct salvage compared with reperfusion alone. The Hemopump appears to provide superior left ventricular systolic and diastolic unloading compared with intra-aortic counterpulsation in a canine model.
引用
收藏
页码:1152 / 1159
页数:8
相关论文
共 19 条
[1]  
ALLEN BS, 1986, J THORAC CARDIOV SUR, V92, P605
[2]   MECHANISM OF MYOCARDIAL STUNNING [J].
BOLLI, R .
CIRCULATION, 1990, 82 (03) :723-738
[3]   RECOVERY OF LEFT-VENTRICULAR SEGMENTAL FUNCTION AFTER LONG-TERM REPERFUSION FOLLOWING TEMPORARY CORONARY-OCCLUSION IN CONSCIOUS DOGS - COMPARISON OF 2-HOUR AND 4-HOUR OCCLUSIONS [J].
BUSH, LR ;
BUJA, LM ;
SAMOWITZ, W ;
RUDE, RE ;
WATHEN, M ;
TILTON, GD ;
WILLERSON, JT .
CIRCULATION RESEARCH, 1983, 53 (02) :248-263
[4]   EFFECT OF SUPEROXIDE-DISMUTASE ON MYOCARDIAL INFARCT SIZE IN THE CANINE HEART AFTER 6 HOURS OF REGIONAL ISCHEMIA AND REPERFUSION - A DEMONSTRATION OF MYOCARDIAL SALVAGE [J].
CHI, L ;
TAMURA, Y ;
HOFF, PT ;
MACHA, M ;
GALLAGHER, KP ;
SCHORK, MA ;
LUCCHESI, BR .
CIRCULATION RESEARCH, 1989, 64 (04) :665-675
[5]   EARLY PHASE ACUTE MYOCARDIAL INFARCT SIZE QUANTIFICATION - VALIDATION OF THE TRIPHENYL TETRAZOLIUM CHLORIDE TISSUE ENZYME STAINING TECHNIQUE [J].
FISHBEIN, MC ;
MEERBAUM, S ;
RIT, J ;
LANDO, U ;
KANMATSUSE, K ;
MERCIER, JC ;
CORDAY, E ;
GANZ, W .
AMERICAN HEART JOURNAL, 1981, 101 (05) :593-600
[6]   FAILURE OF SUPEROXIDE-DISMUTASE AND CATALASE TO ALTER SIZE OF INFARCTION IN CONSCIOUS DOGS AFTER 3 HOURS OF OCCLUSION FOLLOWED BY REPERFUSION [J].
GALLAGHER, KP ;
BUDA, AJ ;
PACE, D ;
GERREN, RA ;
SHLAFER, M .
CIRCULATION, 1986, 73 (05) :1065-1076
[7]   BLOOD-FLOW MEASUREMENTS WITH RADIONUCLIDE-LABELED PARTICLES [J].
HEYMANN, MA ;
PAYNE, BD ;
HOFFMAN, JIE ;
RUDOLPH, AM .
PROGRESS IN CARDIOVASCULAR DISEASES, 1977, 20 (01) :55-79
[8]   LIMITATION OF MYOCARDIAL INFARCT EXPANSION BY REPERFUSION INDEPENDENT OF MYOCARDIAL SALVAGE [J].
HOCHMAN, JS ;
CHOO, H .
CIRCULATION, 1987, 75 (01) :299-306
[9]  
JENNINGS RB, 1983, CIRCULATION, V68, P25
[10]   EFFECT OF INTRA-AORTIC BALLOON COUNTERPULSATION ON MOTION AND PERFUSION OF ACUTELY ISCHEMIC MYOCARDIUM - EXPERIMENTAL ECHOCARDIOGRAPHIC STUDY [J].
KERBER, RE ;
MARCUS, ML ;
EHRHARDT, J ;
ABBOUD, FM .
CIRCULATION, 1976, 53 (05) :853-859