摘要
The accuracy and reliability of the relationship between arterial and end-tidal carbon dioxide (PETCO(2) and Paco(2)), expressed as the gradient, P(a - ET)CO2, was assessed with 171 comparisons in nine mechanically ventilated trauma patients. The P(a - ET)CO2 was 14 +/- 11 mm Hg. (mean +/- standard deviation.) The positive correlation between Paco(2) (44 +/- 10 mm Hg) and PETCO(2) (30 +/- 10 mm Hg) for the study population (reflected by r = 0.64, p = 0.001; but r(2) = 0.41) indicated statistical significance, but only 40% of the changes reflected a linear relationship. Seventyeight percent of individual patients had significant correlations of Paco(2) and PETCO(2) (p < 0.02 to p < 0.001). Changes in PETCO(2) erroneously predicted the Paco(2) changes in 27% of comparisons with 15% false decreases and 12% false increases. Trends in P(a - ET)coa magnitude are not reliable, and concordant direction changes in PETCO(2) and Paco(2) are not assured.
引用
收藏
页码:317 / 322
页数:6
相关论文
共 29 条
[1]   CHANGES IN PHYSIOLOGICAL DEAD SPACE DURING DELIBERATE HYPOTENSION [J].
ASKROG, VF ;
PENDER, JW ;
ECKENHOFF, JE .
ANESTHESIOLOGY, 1964, 25 (6P1) :744-&
[2]  
BERMUDEZ J, 1987, ANESTH ANALG, V66, P690
[3]   THE CONCEPT OF DEADSPACE WITH SPECIAL REFERENCE TO THE SINGLE BREATH TEST FOR CARBON-DIOXIDE [J].
FLETCHER, R ;
JONSON, B ;
CUMMING, G ;
BREW, J .
BRITISH JOURNAL OF ANAESTHESIA, 1981, 53 (01) :77-88
[4]   DEADSPACE AND THE SINGLE BREATH TEST FOR CARBON-DIOXIDE DURING ANESTHESIA AND ARTIFICIAL-VENTILATION [J].
FLETCHER, R ;
JONSON, B .
BRITISH JOURNAL OF ANAESTHESIA, 1984, 56 (02) :109-119
[5]  
FROM RP, 1988, ANESTH ANALG, V67, P884
[6]   ARTERIAL TO END-EXPIRATORY CARBON-DIOXIDE TENSION GRADIENT IN ACUTE PULMONARY-EMBOLISM AND OTHER CARDIOPULMONARY DISEASES [J].
HATLE, L ;
ROKSETH, R .
CHEST, 1974, 66 (04) :352-357
[7]   END-TIDAL CARBON-DIOXIDE IN CRITICALLY ILL PATIENTS DURING CHANGES IN MECHANICAL VENTILATION [J].
HOFFMAN, RA ;
KRIEGER, BP ;
KRAMER, MR ;
SEGEL, S ;
BIZOUSKY, F ;
GAZEROGLU, H ;
SACKNER, MA .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1989, 140 (05) :1265-1268
[8]   CHANGES IN PETCO2 AND PULMONARY BLOOD-FLOW AFTER BRONCHIAL OCCLUSION IN DOGS [J].
JOHNSON, DH ;
CHANG, PC ;
HURST, TS ;
REYNOLDS, FB ;
LANG, SA ;
MAYERS, I .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1992, 39 (02) :184-191
[9]   DIFFERENCE BETWEEN END-TIDAL AND ARTERIAL PCO2 IN EXERCISE [J].
JONES, NL ;
ROBERTSON, DG ;
KANE, JW .
JOURNAL OF APPLIED PHYSIOLOGY, 1979, 47 (05) :954-960
[10]   RELATIONSHIPS BETWEEN INSPIRED OXYGEN CONCENTRATION AND VENOUS ADMIXTURE DURING NITROUS-OXIDE OXYGEN-HALOTHANE ANESTHESIA [J].
KERR, JH ;
FOEX, P ;
PYBUS, DA .
BRITISH JOURNAL OF ANAESTHESIA, 1985, 57 (12) :1150-1160