DOUBLE WAVELENGTH MEASUREMENT OF 3,3',5,5'-TETRAMETHYLBENZIDINE (TMB) PROVIDES A 3-FOLD ENHANCEMENT OF THE ELISA MEASURING RANGE

被引:28
作者
MADERSBACHER, S [1 ]
BERGER, P [1 ]
机构
[1] AUSTRIAN ACAD SCI,IMMUNOENDOCRINOL RES UNIT,FRITZ PREGL STR 3IV,A-6020 INNSBRUCK,AUSTRIA
关键词
ELISA; ASSAY RANGE; HUMAN CHORIONIC GONADOTROPIN; FREE ALPHA-SUBUNIT; 3,3',5,5'-TETRAMETHYLBENZIDINE; HUMAN FOLLICLE STIMULATING HORMONE;
D O I
10.1016/0022-1759(91)90071-M
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A rapid, simple method is described which permits a three-fold enhancement of the working range of ELISA procedures using TMB as a substrate. This consists of measuring absorbance values at a wavelength away from the absorption maximum of TMB and using a predetermined multiplication factor.
引用
收藏
页码:121 / 124
页数:4
相关论文
共 5 条
  • [1] ANTIGENIC FEATURES OF HUMAN FOLLICLE-STIMULATING HORMONE DELINEATED BY MONOCLONAL-ANTIBODIES AND CONSTRUCTION OF AN IMMUNORADIOMOMETRIC ASSAY
    BERGER, P
    PANMOUNG, W
    KHASCHABI, D
    MAYREGGER, B
    WICK, G
    [J]. ENDOCRINOLOGY, 1988, 123 (05) : 2351 - 2359
  • [2] MONOCLONAL-ANTIBODIES AGAINST THE FREE SUBUNITS OF HUMAN CHORIONIC-GONADOTROPIN
    BERGER, P
    KLIEBER, R
    PANMOUNG, W
    MADERSBACHER, S
    WOLF, H
    WICK, G
    [J]. JOURNAL OF ENDOCRINOLOGY, 1990, 125 (02) : 301 - 309
  • [3] Bos E S, 1981, J Immunoassay, V2, P187, DOI 10.1080/15321818108056977
  • [4] Nakane M. B., 1978, IMMUNOFLUORESCENCE R, P215
  • [5] TIJSSEN P, 1985, LABORATORY TECHNIQUE, V15