POTENTIOMETRIC BIOSENSOR IN DOUBLE MATRIX MEMBRANE TECHNOLOGY

被引:15
作者
EGGENSTEIN, C
BORCHARDT, M
DUMSCHAT, C
GRUNDIG, B
CAMMANN, K
SPENER, F
KNOLL, M
机构
[1] Institut für Chemo- und Biosensorik Münster, D 48149 Münster
关键词
POTENTIOMETRIC BIOSENSOR; UREA SENSOR; DISPOSABLE BIOSENSOR;
D O I
10.1016/0956-5663(95)96935-R
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A throw-away urea sensor is described. A pH-sensitive disposable sensor constructed using double matrix membrane technology was used as a transducer. Double matrix membrane technology describes a process where an ion-sensitive polymer matrix membrane is formed in the presence of an additional inert matrix for to improve the reproducibility of sensor production. In this case one-side silver-coated filter paper was used as the inert matrix which was encapsulated by a heat sealing film. A pH-sensitive membrane was casted into this filter paper and on the surface of this membrane a urease containing layer was deposited. The sensor responded to between 10(-2) mol/l and 10(-4) mol/l urea, the response slope was 32.7 mV/decade. The concentration range in which the sensor responds is important for clinical analysis. The main advantages of this pH-based urea electrode include miniature size as well as simple and low cost fabrication in a batch process.
引用
收藏
页码:595 / 600
页数:6
相关论文
共 12 条
  • [1] BORCHARDT M, 1994, IN PRESS SENSORS ACT
  • [2] DOUBLE MATRIX MEMBRANES FOR POTENTIOMETRIC CATION-SELECTIVE DISPOSABLE SENSORS
    DUMSCHAT, C
    BORCHARDT, M
    DIEKMANN, C
    HEPKE, J
    CAMMANN, K
    KNOLL, M
    [J]. FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1994, 348 (8-9): : 553 - 555
  • [3] MICRO-ENZYME FIELD-EFFECT TRANSISTOR SENSOR USING DIRECT COVALENT BONDING OF UREASE
    GARDIES, F
    JAFFREZICRENAULT, N
    MARTELET, C
    PERROT, H
    VALLETON, JM
    ALEGRET, S
    [J]. ANALYTICA CHIMICA ACTA, 1990, 231 (02) : 305 - 308
  • [4] DETERMINATION OF UREA IN BLOOD AND URINE WITH A UREA-SENSITIVE ELECTRODE
    GUILBAULT, GG
    HRABANKOVA, E
    [J]. ANALYTICA CHIMICA ACTA, 1970, 52 (02) : 287 - +
  • [5] SPECIFIC ENZYME ELECTRODE FOR UREA
    GUILBAULT, GG
    TARP, M
    [J]. ANALYTICA CHIMICA ACTA, 1974, 73 (02) : 355 - 365
  • [6] ENZYME SENSOR SYSTEM FOR POTENTIOMETRIC UREA DETERMINATION IN SERUM AND MILK
    HAMANN, H
    KUHN, M
    BOTTCHER, N
    SCHELLER, F
    [J]. JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1986, 209 (01): : 69 - 76
  • [7] MICROFIBER MATRIX-SUPPORTED ION-SELECTIVE PVC MEMBRANES
    KNOLL, M
    CAMMANN, K
    DUMSCHAT, C
    BORCHARDT, M
    HOGG, G
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 1994, 20 (01) : 1 - 5
  • [8] KNOLL M, 1991, Patent No. 372611
  • [9] KUAN SS, 1987, BIOSENSORS
  • [10] RICK W, 1990, KLINISCHE CHEM MIKRO