A LOW FLOW OPEN CIRCLE SYSTEM FOR ANESTHESIA .1. LABORATORY EVALUATION AT NORMAL AND HIGH-FREQUENCIES

被引:6
作者
CHAKRABARTI, MK
STACEY, RGW
HOLDCROFT, A
WHITWAM, JG
机构
[1] Department of Anaesthetics, Royal Postgraduate Medical School, Hammersmith Hospital, London, W12, Du Cane Road
关键词
closed systems; Equipment; ventilators;
D O I
10.1111/j.1365-2044.1990.tb14741.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
A circle breathing system was connected by deadspace tubing to an open system valveless ventilator. The minimum volume of this tubing, required to prevent dilution of anaesthetic gas in the breathing system by the driving gas of the ventilator, was determined at frequencies of 15, 30, 60, 100 and 150 breaths/minute, with tidal volumes that ranged from 100 to 1100 ml and a fresh gas supply to the circle system of 1, 2 and 4 litres/minute. At 15 breaths/minute, tidal volumes equal to or less than the deadspace volume could be used safely without any mixing with the ventilator driving gas, when a fresh gas flow of 2 litres/minute or above is supplied to the circle system. At 1 litre/minute of fresh gas flow, mixing occurred at tidal volumes less than the deadspace volume. Mixing of gas occurred in the system at frequencies greater than 30 breaths/minute even when the tidal volume was much less than the deadspace volume. However, at high frequencies of ventilation, since the tidal volume requirement decreases, deadspace tubing with a safe internal volume, that is greater than 600 ml, may be used up to 100 breaths/minute. Copyright © 1990, Wiley Blackwell. All rights reserved
引用
收藏
页码:314 / 317
页数:4
相关论文
共 5 条
[1]   A NEW VALVELESS ALL-PURPOSE VENTILATOR - DESCRIPTION AND LABORATORY EVALUATION [J].
CHAKRABARTI, MK ;
WHITWAM, JG .
BRITISH JOURNAL OF ANAESTHESIA, 1983, 55 (10) :1005-1015
[2]   RELATIONSHIP BETWEEN FREQUENCY OF VENTILATION, AIRWAYS AND PULMONARY-ARTERY PRESSURES, CARDIAC-OUTPUT AND TRACHEAL TUBE DEADSPACE [J].
CHAKRABARTI, MK ;
GROUNDS, RM ;
SWENZEN, GO ;
WHITWAM, JG .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1986, 30 (08) :678-684
[3]   METHOD OF CONTROLLED VENTILATION WITH A CIRCLE-ABSORBER BREATHING SYSTEM [J].
JEAL, DE .
BRITISH JOURNAL OF ANAESTHESIA, 1977, 49 (03) :273-276
[4]   FACTORS AFFECTING CARBON-DIOXIDE HOMEOSTASIS DURING CONTROLLED VENTILATION WITH CIRCLE SYSTEMS [J].
SCHOONBEE, CG ;
CONWAY, CM .
BRITISH JOURNAL OF ANAESTHESIA, 1981, 53 (05) :471-477
[5]   A NEW VALVELESS ALL-PURPOSE VENTILATOR - CLINICAL-EVALUATION [J].
WHITWAM, JG ;
CHAKRABARTI, MK ;
KONARZEWSKI, WH ;
ASKITOPOULOU, H .
BRITISH JOURNAL OF ANAESTHESIA, 1983, 55 (10) :1017-1023