A STUDY OF DATA LOSS AND MISPOSITIONING DUE TO PILEUP IN 2-D DETECTORS IN PET

被引:46
作者
GERMANO, G [1 ]
HOFFMAN, EJ [1 ]
机构
[1] UNIV CALIF LOS ANGELES,NUCL MED LAB,LOS ANGELES,CA 90024
基金
美国国家卫生研究院;
关键词
D O I
10.1109/23.106696
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The effect of event pileup was investigated with a PET system which employs a 2-D detector-array system. Event pileup caused losses of resolution of 9%, 19.8% and 32% of the FWHM for 10, 20 and 30 mCi of positron emitter in the field of view (FOV), respectively. 9.4%, 18.7% and 28.1% of the data from the outer planes of the module were incorrectly placed in the inner planes for 5, 10, and 15 mCi of positron emitter in the FOV, respectively. A simple model of the process was formulated and was found to be adequate to describe the major features of event pileup for a number of source configurations and a wide range of activities. Copyright © 1990 by The Institute of Electrical and Electronics Engineers, Inc.
引用
收藏
页码:671 / 675
页数:5
相关论文
共 24 条
[1]  
[Anonymous], TECHNICAL SPECIFICAT
[2]  
ARNOLD JE, 1974, J NUCL MED, V15, P412
[3]  
BRENNER R, 1982, IEEE T NUCL SCI, V29, P1
[4]   HIGH-RESOLUTION AND HIGH-SPEED POSITRON EMISSION TOMOGRAPHY DATA ACQUISITION [J].
BURGISS, SG ;
BYARS, LG ;
JONES, WF ;
CASEY, ME .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1986, 33 (01) :489-491
[5]  
CARSON RE, 1981, 5TH P ANN S COMP APP
[6]   A MULTICRYSTAL 2-DIMENSIONAL BGO DETECTOR SYSTEM FOR POSITRON EMISSION TOMOGRAPHY [J].
CASEY, ME ;
NUTT, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1986, 33 (01) :460-463
[7]   NOISE AND PILEUP SUPPRESSION BY DIGITAL SIGNAL-PROCESSING [J].
CHRIEN, RE ;
SUTTER, RJ .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1986, 249 (2-3) :421-425
[8]   AN EVALUATION OF A TWO-DIMENSIONAL ARRAY DETECTOR FOR HIGH-RESOLUTION PET [J].
DAHLBOM, M ;
HOFFMAN, EJ .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1988, 7 (04) :264-272
[9]   A STUDY OF THE POSSIBILITY OF USING MULTI-SLICE PET SYSTEMS FOR 3D IMAGING [J].
DAHLBOM, M ;
ERIKSSON, L ;
ROSENQVIST, G ;
BOHM, C .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1989, 36 (01) :1066-1071
[10]  
DAHLBOM M, 1987, THESIS U CALIFORNIA