DETERMINATION OF SUBNANOGRAM PER CUBIC METER CONCENTRATIONS OF METALS IN THE AIR OF A TRACE-METAL CLEAN ROOM BY IMPACTION GRAPHITE-FURNACE ATOMIC-ABSORPTION AND LASER-EXCITED ATOMIC FLUORESCENCE SPECTROMETRY

被引:26
作者
LIANG, ZW
WEI, GT
IRWIN, RL
WALTON, AP
MICHEL, RG
机构
[1] UNIV CONNECTICUT,DEPT CHEM,215 GLENBROOK RD,STORRS,CT 06269
[2] UNIV LOWELL,DEPT CHEM,LOWELL,MA 01854
关键词
D O I
10.1021/ac00213a020
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Air, drawn by vacuum through a jet, was impacted against the inside surface of an atomic absorption graphite electrothermal atomizer (ETA). The amounts of the particles thus collected were determined at the ng m-3 level by graphite furnace atomic absorption or at the pg m-3 level by laser excited atomic fluorescence. The overall reproducibility of two sets of measurements, made 7 months apart, was 23%, with no significant difference between the two sets of data, based on Student's “t” test at the 95% confidence level. Short-term reproducibility varied from 13% to 34% depending upon the air concentration of the metal. The method shows promise for monitoring long-term effectiveness of the filtering systems in trace metal clean rooms. It was not possible to test for accuracy, due to the low concentrations involved, but accuracy was expected to be within a factor of 2 or 3 of the actual value, based on theoretical aspects of impaction. © 1990, American Chemical Society. All rights reserved.