A COMPUTER-ASSISTED METAL ANALYZER USING FLOW-INJECTION COUPLED WITH DIRECT-CURRENT PLASMA - OPTICAL-EMISSION SPECTROSCOPY

被引:4
作者
BRENNAN, MC [1 ]
SIMONS, RA [1 ]
SVEHLA, G [1 ]
STOCKWELL, PB [1 ]
机构
[1] PS ANALYT LTD, SEVENOAKS TN15 6QY, KENT, ENGLAND
来源
JOURNAL OF AUTOMATIC CHEMISTRY | 1990年 / 12卷 / 05期
关键词
D O I
10.1155/S1463924690000232
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
An automated routine method was developed for the determination of trace metals using a flow injection system into a direct current plasma spectrometer. The characteristic emission in the ‘Spectra-Metrics’ three-electrode jet is channelled through an echelle grating to the detector of the instrument. The emission signal is continuously recorded, and, after digitalization, calculated and recorded by a Touchstone software package and interface card. The software has facilities for recording peaks, generating calibration curves, calculating and printing results, enhancing line sensitivities, monitoring signal response and for filter noise, recalibration, background correction, status monitoring and general diagnostics. Up to 70 elements can be determined both in aqueous and non-aqueous media. © 1990, Hindawi Publishing Corporation. All rights reserved.
引用
收藏
页码:183 / 188
页数:6
相关论文
共 21 条
[1]  
BRENNAN MC, 1989, Z ANAL CHEM, V335, P893
[2]   A LOW CONSUMPTION THERMOSPRAY NEBULIZER WITH A FUSED-SILICA VAPORIZER FOR INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
ELGERSMA, JW ;
MAESSEN, FJMJ ;
NIESSEN, WMA .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1986, 41 (11) :1217-1220
[3]  
FANG Z, 1986, CHINESE J ANAL CHEM, V14, P549
[4]   MICROLITER SAMPLE INTRODUCTION FOR ICP-AES [J].
FASKE, AJ ;
SNABLE, KR ;
BOORN, AW ;
BROWNER, RF .
APPLIED SPECTROSCOPY, 1985, 39 (03) :542-545
[5]  
Gallego M, 1985, ATOM SPECTROSC, V6, P16
[6]  
GREENFIELD S, 1981, RES DEV, V23, P140
[8]   SENSITIVITY ENHANCEMENTS FOR FLOW-INJECTION ANALYSIS INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY USING AN ONLINE PRECONCENTRATING ION-EXCHANGE COLUMN [J].
HARTENSTEIN, SD ;
RUZICKA, J ;
CHRISTIAN, GD .
ANALYTICAL CHEMISTRY, 1985, 57 (01) :21-25
[9]   STANDARD ADDITION METHOD IN FLOW-INJECTION ANALYSIS WITH INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
ISRAEL, Y ;
BARNES, RM .
ANALYTICAL CHEMISTRY, 1984, 56 (07) :1188-1192
[10]   FLOW-INJECTION SYSTEMS WITH INDUCTIVELY-COUPLED ARGON PLASMA ATOMIC EMISSION-SPECTROMETRY .1. FUNDAMENTAL CONSIDERATIONS [J].
JACINTHO, AO ;
ZAGATTO, EAG ;
BERGAMIN, H ;
KRUG, FJ ;
REIS, BF ;
BRUNS, RE ;
KOWALSKI, BR .
ANALYTICA CHIMICA ACTA, 1981, 130 (02) :243-255