QUALITY ASSURANCE IN RADIOTHERAPY BY INVIVO DOSIMETRY .1. ENTRANCE DOSE MEASUREMENTS, A RELIABLE PROCEDURE

被引:112
作者
LEUNENS, G
VANDAM, J
DUTREIX, A
VANDERSCHUEREN, E
机构
[1] Department of Radiotherapy, University Hospital St. Rafael, Leuven
关键词
In vivo dosimetry; Quality assurance; Silicon diodes;
D O I
10.1016/0167-8140(90)90102-3
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Entrance dose measurements were performed with semiconductor detectors on patients treated for head and neck and brain tumors with a 6 MV X-ray beam. A total number of 554 treatment set-ups were measured. The results showed a gaussian distribution with a mean value of 97.8% and a standard deviation of 2.8%. A systematic error of 2.2% on the mean value was shown to be due to a systematic deficiency in the algorithm used in the planning system and to a systematic error in the application of the dosimetry protocol. Two treatment techniques were identified leading to an erroneous dose delivery. Finally, large deviations (more than 2 S.D.) of the measured dose from the expected dose were detected in 3% of the measured treatment set-ups, the sources of the errors could in all cases be identified and eliminated in the further treatment sessions. This study demonstrated the reliability of the use of semiconductor detectors for in vivo dosimetry and its usefulness as part of a departmental quality assurance program. © 1990.
引用
收藏
页码:141 / 151
页数:11
相关论文
共 20 条
[1]   DOSIMETRIC PRECISION REQUIREMENTS IN RADIATION-THERAPY [J].
BRAHME, A .
ACTA RADIOLOGICA ONCOLOGY, 1984, 23 (05) :379-391
[2]  
Cameron JR, 1968, THERMOLUMINESCENT DO
[3]  
CHAVAUDRA J, 1976, J RADIOL ELEC MED N, V57, P435
[4]   WHEN AND HOW CAN WE IMPROVE PRECISION IN RADIOTHERAPY [J].
DUTREIX, A .
RADIOTHERAPY AND ONCOLOGY, 1984, 2 (04) :275-292
[5]   CALCULATION OF THE UNCERTAINTY IN THE DOSE DELIVERED DURING RADIATION-THERAPY [J].
GOITEIN, M .
MEDICAL PHYSICS, 1985, 12 (05) :608-612
[6]   EVALUATION OF TEMPERATURE EFFECTS IN P-TYPE SILICON DETECTORS [J].
GRUSELL, E ;
RIKNER, G .
PHYSICS IN MEDICINE AND BIOLOGY, 1986, 31 (05) :527-534
[7]  
HERRING DF, 1971, 3RD P INT C COMP RAD
[8]  
JOHANSSON KA, 1982, THESIS U GOTEBORG SW
[9]  
LAGRANGE JL, 1988, CONTRIBUTION OPTIMIS
[10]  
LINDSKOUG B, 1974, THESIS U GOTEBORG SW