THE EFFECT OF PATIENT MOTION ON DOSE UNCERTAINTY IN CHARGED-PARTICLE IRRADIATION FOR LESIONS ENCIRCLING THE BRAIN-STEM OR SPINAL-CORD

被引:7
作者
DAFTARI, I
PETTI, PL
COLLIER, JM
CASTRO, JR
PITLUCK, S
机构
[1] Division of Research Medicine and Radiation Biophysics University of California, Lawrence Berkeley, Laboratory
关键词
CHARGED-PARTICLE THERAPY; LOCALIZATION ACCURACY; PATIENT MOTION; TREATMENT PLANNING;
D O I
10.1118/1.596619
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A specialized charged-particle radiotherapy technique developed at Lawrence Berkeley Laboratory (LBL) is applied to patients with lesions abutting or surrounding the spinal cord or brain stem. This technique divides the target into two parts, one partially surrounding the critical structure (brain stem or spinal cord) and a second excluding the critical structure and abutting the first portion of the target. Compensators are used to conform the dose distribution to the distal surface of the target. This technique represents a novel approach in treating unresectable or residual tumors surrounding the spinal cord or brain stem. Since the placement of the patient with respect to beam-shaping devices is critical for divided-target treatments, a method for calculating dose distributions reflecting random patient motion is proposed, and the effects of random patient motion are studied for two divided-target patient examples. Dose-volume histograms and a normal-tissue complication probability model are used in this analysis. For the patients considered in this study, the normal-tissue-complication probability model predicts that random patient motion less than or equal to 0.2 cm is tolerable in terms of spinal cord complications.
引用
收藏
页码:1105 / 1115
页数:11
相关论文
共 23 条
[1]   RADIATION MYELITIS OF THE CERVICAL SPINAL CORD [J].
BODEN, G .
BRITISH JOURNAL OF RADIOLOGY, 1948, 21 (249) :464-469
[2]   CHARGED-PARTICLE RADIOTHERAPY FOR LESIONS ENCIRCLING THE BRAIN-STEM OR SPINAL-CORD [J].
CASTRO, JR ;
COLLIER, JM ;
PETTI, PL ;
NOWAKOWSKI, V ;
CHEN, GTY ;
LYMAN, JT ;
LINSTADT, D ;
GAUGER, G ;
GUTIN, P ;
DECKER, M ;
PHILLIPS, TL ;
BAKEN, K .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1989, 17 (03) :477-484
[3]  
CHEN CTY, 1983, ADV MED PHYSICS MONO, V9, P514
[4]  
Chen G. T. Y., 1985, Proceedings of the Sixth Annual Conference and Exposition: Computer Graphics '85. Vol. III, Technical Sessions, P171
[5]   TREATMENT PLANNING FOR HEAVY-ION RADIOTHERAPY [J].
CHEN, GTY ;
SINGH, RP ;
CASTRO, JR ;
LYMAN, JT ;
QUIVEY, JM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1979, 5 (10) :1809-1819
[6]  
CHEN GTY, 1986, MED PHYSICS MONOGRAP, V15, P595
[7]   AN ISO-EFFECT TABLE FOR RADIATION TOLERANCE OF THE HUMAN SPINAL-CORD [J].
COHEN, L ;
CREDITOR, M .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1981, 7 (07) :961-966
[8]  
COLLIER JM, 1990, MED PHYS, V17, P552
[9]   MULTIDIMENSIONAL TREATMENT PLANNING .2. BEAM EYE-VIEW, BACK PROJECTION, AND PROJECTION THROUGH CT SECTIONS [J].
GOITEIN, M ;
ABRAMS, M ;
ROWELL, D ;
POLLARI, H ;
WILES, J .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1983, 9 (06) :789-797
[10]   COMPENSATION FOR INHOMOGENEITIES IN CHARGED-PARTICLE RADIOTHERAPY USING COMPUTED TOMOGRAPHY [J].
GOITEIN, M .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1978, 4 (5-6) :499-508