SNARE REMOVAL OF A TELECTRONICS ACCUFIX ATRIAL J RETENTION WIRE

被引:11
作者
LLOYD, MA
HAYES, DL
STANSON, AW
HOLMES, DR
机构
[1] MAYO CLIN & MAYO FDN,DIV CARDIOVASC DIS,ROCHESTER,MN 55905
[2] MAYO CLIN & MAYO FDN,DIV INTERNAL MED,ROCHESTER,MN 55905
[3] MAYO CLIN & MAYO FDN,DEPT DIAGNOST RADIOL,ROCHESTER,MN 55905
关键词
D O I
10.4065/70.4.376
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A voluntary recall of the Telectronics Accufix models 330-801 and 329-701 ''J''-shaped atrial pacemaker leads has been issued because of fracture and extrusion of the J-shaped retention wire. Such extrusion can result in laceration of the atrium or surrounding vascular structures, Herein we describe a patient with a fractured and extruded retention wire that was snared and removed percutaneously; the atrial lead was left intact, and pacing function was satisfactory. This unique procedure can be used to avoid the morbidity associated with percutaneous lead extraction or thoracotomy and to prevent potential dislodgment and embolization of the retention wire during lead extraction.
引用
收藏
页码:376 / 379
页数:4
相关论文
共 5 条
  • [1] BRINKER JA, 1991, PACE, V14, P639
  • [2] DENNIS JW, 1995, TELECTRONICS PA 1203, P12
  • [3] BIPOLAR TINED POLYURETHANE VENTRICULAR LEAD - A 4-YEAR EXPERIENCE
    HAYES, DL
    HOLMES, DR
    MERIDETH, J
    OSBORN, MJ
    VLIETSTRA, RE
    NEUBAUER, SA
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1985, 8 (02): : 192 - 196
  • [4] A MULTICENTER EXPERIENCE WITH A BIPOLAR TINED POLYURETHANE VENTRICULAR LEAD
    HAYES, DL
    GRAHAM, KJ
    IRWIN, M
    VIDAILLET, H
    DISLER, G
    SWEESY, M
    OSBORN, MJ
    SUMAN, VJ
    NEUBAUER, SA
    SEEBANDT, M
    KALLINEN, L
    CROWSON, CS
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (07): : 1033 - 1039
  • [5] 5-YEARS EXPERIENCE WITH INTRAVASCULAR LEAD EXTRACTION
    SMITH, HJ
    FEARNOT, NE
    BYRD, CL
    WILKOFF, BL
    LOVE, CJ
    SELLERS, TD