COMPARISON OF AUTOMATIC OSCILLOMETRIC ARTERIAL-PRESSURE MEASUREMENT WITH CONVENTIONAL AUSCULTATORY MEASUREMENT IN THE LABOR WARD

被引:28
作者
HASAN, MA [1 ]
THOMAS, TA [1 ]
PRYSROBERTS, C [1 ]
机构
[1] BRISTOL MATERN HOSP,DEPT ANAESTHESIA,BRISTOL BS2 8EG,ENGLAND
关键词
ARTERIAL PRESSURE; MEASUREMENT; EQUIPMENT; MANOMETERS; PREGNANCY;
D O I
10.1093/bja/70.2.141
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We have compared two non-invasive methods of arterial pressure (AP) measurement used in labour wards: an automatic oscillometric measurement obtained by Dinamap 1846, and a conventional auscultatory measurement obtained by midwives. A total of 369 AP measurements were recorded, involving 28 normotensive and hypertensive pregnant women during labour, with or without extra-dural analgesia. Compared with the midwife group, the Dinamap group had a greater systolic AP, by 2.7 mm Hg (P < 0.01) and smaller diastolic AP, by 9.8 mm Hg (P < 0.01). The correlations between the two methods were highly significant, but the limits of agreement were relatively wide for both systolic and diastolic AP measurements. We conclude that a clinically important difference exists in diastolic AP measurements. Dinamap diastolic AP must be corrected using a regression equation, or simply by adding 10 mm Hg, before being compared with the available normal and hypertensive AP values.
引用
收藏
页码:141 / 144
页数:4
相关论文
共 14 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   THE CLASSIFICATION AND DEFINITION OF THE HYPERTENSIVE DISORDERS OF PREGNANCY [J].
DAVEY, DA ;
MACGILLIVRAY, I .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1988, 158 (04) :892-898
[3]  
FRIEDMAN EA, 1977, PREGNANCY HYPERTENS, P238
[4]  
FRIESEN RH, 1981, ANESTH ANALG, V60, P742
[5]   ACCURACY OF 4 INDIRECT METHODS OF BLOOD-PRESSURE MEASUREMENT, WITH HEMODYNAMIC CORRELATIONS [J].
GRAVLEE, GP ;
BROCKSCHMIDT, JK .
JOURNAL OF CLINICAL MONITORING, 1990, 6 (04) :284-298
[6]   AN ASSESSMENT OF THE DINAMAP 845 [J].
HUTTON, P ;
DYE, J ;
PRYSROBERTS, C .
ANAESTHESIA, 1984, 39 (03) :261-267
[7]   AN AUTOMATED OSCILLOMETRIC TECHNIQUE FOR ESTIMATING MEAN ARTERIAL-PRESSURE IN CRITICALLY ILL NEWBORNS [J].
KIMBLE, KJ ;
DARNALL, RA ;
YELDERMAN, M ;
ARIAGNO, RL ;
REAM, AK .
ANESTHESIOLOGY, 1981, 54 (05) :423-425
[8]  
MACGILLIVRAY I, 1969, CLIN SCI, V37, P395
[9]  
MARX GF, 1991, ANESTH ANALG, V72, P562
[10]   NON-INVASIVE AUTOMATIC-DETERMINATION OF MEAN ARTERIAL-PRESSURE [J].
RAMSEY, M .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1979, 17 (01) :11-18