ACL substitutes made of braided or plied purified collagen fibers and cross-linked with hexamethylenediisocyanate were implanted into a total of 14 adult goats to achieve resorption within 8 to 10 months. Two types of collagen fiber prostheses differing in degree of collagen purification were tested. The implants were harvested 2 to 11 months postimplantation, tested for mechanical strength, and evaluated by morphological methods. In the first group (n = 5), the less purified and less cross-linked collagen fiber ACL implant induced fast connective tissue ingrowth. At 6 months postimplantation, 40 to 60% of the collagen implant was resorbed. No studies on breaking strength were done in this group. In the second group, highly purified and more cross-linked ACL implants were less infiltrated by cells and were resorbed only by 10 to 20%. Still, the breaking strength was decreased to 10% of the original implant strength, In the second group, the fixation of the ACL implant in the bone tunnel with a bone wedge was insufficient (n = 6); however, additional fixation with metal screws was successful (n = 3). We conclude that cross-linked collagen fibers alone cannot be used as a safe ACL substitute as they quickly lose mechanical strength despite limited biodegradation.