SEMIEMPIRICAL MODEL FOR GENERATING TUNGSTEN TARGET X-RAY-SPECTRA

被引:388
作者
TUCKER, DM [1 ]
BARNES, GT [1 ]
CHAKRABORTY, DP [1 ]
机构
[1] UNIV ALABAMA,UNIV ALABAMA HOSP & CLIN,DEPT RADIOL,BIRMINGHAM,AL 35294
关键词
D O I
10.1118/1.596709
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A semiempirical model for generating tungsten target x-ray spectra is presented. This model extends earlier work in two significant areas. First, both bremsstrahlung and characteristic x-ray production are assumed to occur at varying depths within the target. Second, optimal parameters for the model were determined from experimental spectra utilizing nonlinear least-squares techniques. As a result, good agreement is obtained between calculated and measured x-ray tube spectra and output for different target angles and a wide range of x-ray tube potentials. Such is not the case with previously published models.
引用
收藏
页码:211 / 218
页数:8
相关论文
共 26 条
  • [1] Attix F. H., 1986, INTRO RADIOLOGICAL P, P560
  • [2] BERGER MJ, 1983, NBSIR822550A, P102
  • [3] Bevington P.R., 1969, DATA REDUCTION ERROR
  • [4] BICHSEL H, 1968, RAD DOSIMETRY, V1, P189
  • [5] COMPUTATION OF BREMSSTRAHLUNG X-RAY-SPECTRA AND COMPARISON WITH SPECTRA MEASURED WITH A GE(LI) DETECTOR
    BIRCH, R
    MARSHALL, M
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1979, 24 (03) : 505 - 517
  • [6] BIRCH R, 1979, CATALOGUE SPECTRAL D
  • [7] CULLEN DE, 1989, UCRL50400 NAT TECHN, V6
  • [8] DYSON NA, 1973, XRAYS ATOMIC NUCLEAR, P107
  • [9] DYSON NA, 1973, XRAYS ATOMIC NUCLEAR, P98
  • [10] EVANS RD, 1955, ATOMIC NUCLEUS, P602