VALIDATION OF COMPUTER ANALYZED POLYGRAPHIC PATTERNS DURING DROWSINESS AND SLEEP ONSET

被引:23
作者
HASAN, J
HIRVONEN, K
VARRI, A
HAKKINEN, V
LOULA, P
机构
[1] TAMPERE UNIV HOSP,DEPT CLIN NEUROPHYSIOL,POB 2000,SF-33521 TAMPERE,FINLAND
[2] TAMPERE UNIV TECHNOL,SF-33101 TAMPERE 10,FINLAND
[3] RES CTR FINLAND,MED ENGN LAB,TAMPERE,FINLAND
来源
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY | 1993年 / 87卷 / 03期
基金
芬兰科学院;
关键词
AUTOMATIC ANALYSIS; DROWSINESS; SLEEP; VIGILANCE; ADAPTIVE SEGMENTATION;
D O I
10.1016/0013-4694(93)90118-F
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A computer system for the automatic analysis of polygraphic records was validated. Records from 9 subjects made during routine MSLT tests were analysed both by two preliminary and one consensus scorer and by a computer system. Special attention was paid to the analysis of drowsiness periods. Therefore a classification system including 7 stages, three for wakefulness and movement, one for drowsiness and three for the sleep stages Sl, S2 and SREM was used. Adaptive segmentation was used to divide the records into short segments of variable length (mean 1.6 sec, range 0.5-13.7 sec). The agreements between the computer and visual scores were relatively good for 5 subjects having a prominent occipital alpha activity during wakefulness (range 70-79%) but less promising (range 64-70%) for the other 4 subjects with ''poor'' occipital alpha activity. The values obtained corresponded to the inter-scorer agreements. Most of the discrepancies were between adjacent stages. At times in the presence of strongly fluctuating EEG amplitudes and especially with the ''low-alpha'' subjects it was very difficult to determine exactly even by visual scoring when, for instance, drowsiness became sleep. It is concluded that the reliability of the system is sufficient for practical purposes especially if critical parts of the records are visually reexamined. It was found to be difficult to define unambiguous scoring criteria for subjects with poorly defined EEG rhythms giving insufficient landmarks for stage determination.
引用
收藏
页码:117 / 127
页数:11
相关论文
共 55 条
[1]   SLEEPINESS AS A CONSEQUENCE OF SHIFT WORK [J].
AKERSTEDT, T .
SLEEP, 1988, 11 (01) :17-34
[2]   AUTOMATIC ADAPTIVE SEGMENTATION OF CLINICAL EEGS [J].
BARLOW, JS ;
CREUTZFELDT, OD ;
MICHAEL, D ;
HOUCHIN, J ;
EPELBAUM, H .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1981, 51 (05) :512-525
[3]  
BELAYAVIN A, 1987, ELECTROENCEPHALOGR C, V66, P137
[4]  
BENTE D, 1965, THESIS ERLANGEN
[5]   FEATURE EXTRACTION FROM ELECTROENCEPHALOGRAM BY ADAPTIVE SEGMENTATION [J].
BODENSTEIN, G ;
PRAETORIUS, HM .
PROCEEDINGS OF THE IEEE, 1977, 65 (05) :642-652
[6]   AMBULATORY 24 HOUR SLEEP-WAKE MONITORING IN NARCOLEPSY-CATAPLEXY COMPARED TO MATCHED CONTROLS [J].
BROUGHTON, R ;
DUNHAM, W ;
NEWMAN, J ;
LUTLEY, K ;
DUSCHESNE, P ;
RIVERS, M .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1988, 70 (06) :473-481
[7]   GUIDELINES FOR THE MULTIPLE SLEEP LATENCY TEST (MSLT) - A STANDARD MEASURE OF SLEEPINESS [J].
CARSKADON, MA ;
DEMENT, WC ;
MITLER, MM ;
ROTH, T ;
WESTBROOK, PR ;
KEENAN, S .
SLEEP, 1986, 9 (04) :519-524
[8]   A COEFFICIENT OF AGREEMENT FOR NOMINAL SCALES [J].
COHEN, J .
EDUCATIONAL AND PSYCHOLOGICAL MEASUREMENT, 1960, 20 (01) :37-46
[9]   COMPUTERIZED EEG PATTERN-CLASSIFICATION BY ADAPTIVE SEGMENTATION AND PROBABILITY DENSITY-FUNCTION CLASSIFICATION - CLINICAL-EVALUATION [J].
CREUTZFELDT, OD ;
BODENSTEIN, G ;
BARLOW, JS .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1985, 60 (05) :373-393
[10]   ALPHA AND THETA EEG IN VIGILANCE [J].
DANIEL, RS .
PERCEPTUAL AND MOTOR SKILLS, 1967, 25 (03) :697-&