HETEROGENEITY IN GENTAMICIN CLEARANCE BETWEEN HIGH-EFFICIENCY HEMODIALYZERS

被引:18
作者
AGARWAL, R
CRONIN, RE
机构
[1] VET AFFAIRS MED CTR,DIV RENAL,DALLAS,TX 75216
[2] UNIV TEXAS,SW MED CTR,DEPT INTERNAL MED,DIV NEPHROL,DALLAS,TX
关键词
GENTAMICIN; HEMODIALYSIS; ELIMINATION RATE CONSTANT; HIGH-EFFICIENCY DIALYSIS; THERAPEUTIC DRUG MONITORING;
D O I
10.1016/S0272-6386(12)80811-0
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Gentamicin is frequently prescribed for treatment of gram-negative infections in patients on dialysis. The aminoglycoside is renally excreted; therefore, dosage modifications are required in patients on hemodialysis. The introduction of short-duration dialysis using high-efficiency hemodialyzers requires that the dialysance of gentamicin be re-examined. We recently demonstrated that high-efficiency dialyzers have a superior gentamicin clearance as compared with conventional dialyzers. In this study we examined the dialysance of gentamicin in two high-efficiency dialyzers using a randomized, controlled, cross-over design. Despite a comparable amount of dialysis (Kt/V urea), rate of dialysis (K/ V urea), and ultrafiltration, there were significant differences noted in gentamicin removal between the high-efficiency cuprammonium rayon and polysulfone dialyzers. Furthermore, the amount of dialysis (Kt/V urea) could predict the fractional gentamicin clearance for the TAF 175L dialyzer (Terumo Corp, Somerset, NJ) and may be useful in the initial planning of gentamicin therapy. In addition, since gentamicin is a middle molecule (molecular weight, 500 to 2,000 d), the study suggests a superior middle molecule clearance of cuprammonium rayon dialyzer compared with a highefficiency polysulfone dialyzer. The substantial heterogeneity in gentamicin clearance, even between dialyzers of the same class, emphasizes the importance of monitoring drug levels in hemodialysis patients receiving this drug. © 1994, National Kidney Foundation. All rights reserved.. All rights reserved.
引用
收藏
页码:47 / 51
页数:5
相关论文
共 11 条
[1]   GENTAMICIN CLEARANCE DURING HEMODIALYSIS - A COMPARISON OF HIGH-EFFICIENCY CUPRAMMONIUM RAYON AND CONVENTIONAL CELLULOSE ESTER HEMODIALYZERS [J].
AGARWAL, R ;
TOTO, RD .
AMERICAN JOURNAL OF KIDNEY DISEASES, 1993, 22 (02) :296-299
[2]   REBOUND GENTAMICIN LEVELS AFTER HEMODIALYSIS [J].
BAUER, LA .
THERAPEUTIC DRUG MONITORING, 1982, 4 (01) :99-101
[3]   DOSE OF DIALYSIS - WHAT INDEX [J].
CHARRA, B ;
CALEMARD, E ;
CHAZOT, C ;
TERRAT, JC ;
VANEL, T ;
RUFFET, M ;
LAURENT, G .
BLOOD PURIFICATION, 1992, 10 (01) :13-21
[4]   HEMODIALYZER CLEARANCES OF GENTAMICIN, KANAMYCIN, TOBRAMYCIN, AMIKACIN, ETHAMBUTOL, PROCAINAMIDE, AND FLUCYTOSINE, WITH A TECHNIQUE FOR PLANNING THERAPY [J].
CHRISTOPHER, TG ;
BLAIR, AD ;
FORREY, AW ;
CUTLER, RE .
JOURNAL OF PHARMACOKINETICS AND BIOPHARMACEUTICS, 1976, 4 (05) :427-441
[5]   PHARMACOKINETICS OF GENTAMICIN AND KANAMYCIN DURING HEMODIALYSIS [J].
DANISH, M ;
SCHULTZ, R ;
JUSKO, WJ .
ANTIMICROBIAL AGENTS AND CHEMOTHERAPY, 1974, 6 (06) :841-847
[6]   PREDICTABILITY OF SERUM CONCENTRATIONS OF AMINOGLYCOSIDES AFTER HEMODIALYSIS [J].
DIJKMANS, BAC ;
MATTIE, H .
EUROPEAN JOURNAL OF CLINICAL PHARMACOLOGY, 1987, 33 (02) :179-183
[7]   PREDICTION OF SERUM GENTAMICIN CONCENTRATIONS IN PATIENTS UNDERGOING HEMODIALYSIS [J].
GOETZ, DR ;
PANCORBO, S ;
HOAG, S ;
BLOOM, P .
AMERICAN JOURNAL OF HOSPITAL PHARMACY, 1980, 37 (08) :1077-1083
[8]   CLEARANCE OF GENTAMICIN DURING HEMODIALYSIS - COMPARISON OF 4 ARTIFICIAL-KIDNEYS [J].
HALPREN, BA ;
AXLINE, SG ;
COPLON, NS ;
BROWN, DM .
JOURNAL OF INFECTIOUS DISEASES, 1976, 133 (06) :627-636
[9]   HEMODIALYSIS ELIMINATION RATES AND CLEARANCE OF GENTAMICIN AND TOBRAMYCIN [J].
MATZKE, GR ;
HALSTENSON, CE ;
KEANE, WF .
ANTIMICROBIAL AGENTS AND CHEMOTHERAPY, 1984, 25 (01) :128-130
[10]  
RITSCHEL WA, 1986, HDB BASIC PHARMACOKI, P168