TEMPERATURE AND SAR MEASUREMENTS IN DEEP-BODY HYPERTHERMIA WITH THERMOCOUPLE THERMOMETRY

被引:54
作者
DELEEUW, AAC
CREZEE, J
LAGENDIJK, JJW
机构
[1] Department of Radiotherapy, University Hospital, Heidelberglaan, Utrecht
关键词
REGIONAL EM HYPERTHERMIA; THERMOCOUPLE THERMOMETRY; SAR MEASUREMENT;
D O I
10.3109/02656739309032056
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Multisensor (7-14) thermocouple thermometry is used at our department for temperature measurement with our 'Coaxial TEM' regional hyperthermia system. A special design of the thermometry system with high resolution (0.005-degrees-C) and fast data-acquisition (all channels within 320 ms) together with a pulsed power technique allows assessment of specific absorption rate (SAR) information in patients along catheter tracks. A disadvantage of thermocouple thermometry, EM interference, is almost entirely eliminated by application of absorbing ferrite beads around the probe leads. We investigated the effect of remaining disturbance on the temperature decay after power-off, both experimentally in phantoms and in the clinic, and with numerical simulations. Probe and tissue characteristics influence the response time tau(dist) of the decay of the disturbance. In our clinical practice a normal pulse sequence is 50 s power-on, 10 s power-off: a response time longer than the power-off time results in a deflection of the temperature course at the start. Based on analysis of temperature decays correction of temperature is possible. A double-pulse technique is introduced to provide an initial correction of temperature, and fast information about accuracy. Sometimes disturbance with a relatively long response time occurs, probably due to a bad contact between probe, catheter and/or tissue. Thermocouple thermometry proved to be suitable to measure the SAR along a catheter track. This is used to optimize the SAR distribution by patient positioning before treatment. A clinical example illustrates this.
引用
收藏
页码:685 / 697
页数:13
相关论文
共 23 条
[1]  
ASTRAHAN MA, 1989, STRAHLENTHER ONKOL, V165, P746
[2]   PERIPHERAL NEUROPATHY AND MYONECROSIS FOLLOWING HYPERTHERMIA AND RADIATION-THERAPY FOR RECURRENT PROSTATIC-CANCER - CORRELATION OF DAMAGE WITH PREDICTED SAR PATTERN [J].
BENYOSEF, R ;
SULLIVAN, DM ;
KAPP, DS .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1992, 8 (02) :173-185
[3]  
Bolomey J. C., 1990, METHODS HYPERTHERMIA, P35
[4]  
CARNOCHAN P, 1986, International Journal of Hyperthermia, V2, P1, DOI 10.3109/02656738609019990
[5]   INTERSTITIAL HEATING - EXPERIMENTS IN ARTIFICIALLY PERFUSED BOVINE TONGUES [J].
CREZEE, J ;
MOOIBROEK, J ;
BOS, CK ;
LAGENDIJK, JJW .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (06) :823-833
[6]  
Croft D.R., 1977, HEAT TRANSFER CALCUL
[7]   SPECIFIC ABSORPTION RATE STEERING BY PATIENT POSITIONING IN THE COAXIAL TEM SYSTEM - PHANTOM INVESTIGATION [J].
DELEEUW, AAC ;
MOOIBROEK, J ;
LAGENDIJK, JJW .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1991, 7 (04) :605-611
[8]   DESIGN OF A CLINICAL DEEP-BODY HYPERTHERMIA SYSTEM BASED ON THE COAXIAL TEM APPLICATOR [J].
DELEEUW, AAC ;
LAGENDIJK, JJW .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1987, 3 (05) :413-421
[9]   RTOG QUALITY ASSURANCE GUIDELINES FOR CLINICAL-TRIALS USING HYPERTHERMIA [J].
DEWHIRST, MW ;
PHILLIPS, TL ;
SAMULSKI, TV ;
STAUFFER, P ;
SHRIVASTAVA, P ;
PALIWAL, B ;
PAJAK, T ;
GILLIM, M ;
SAPOZINK, M ;
MYERSON, R ;
WATERMAN, FM ;
SAPARETO, SA ;
CORRY, P ;
CETAS, TC ;
LEEPER, DB ;
FESSENDEN, P ;
KAPP, D ;
OLESON, JR ;
EMAMI, B .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 18 (05) :1249-1259
[10]   THERMAL CONDUCTION ERRORS OF MANGANIN-CONSTANTAN THERMOCOUPLE ARRAYS [J].
DICKINSON, RJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1985, 30 (05) :445-453