MIRAU CORRELATION MICROSCOPE

被引:378
作者
KINO, GS
CHIM, SSC
机构
[1] Stanford University, Hansen Laboratories of Physics, Ginzton Laboratory, Stanford, CA
来源
APPLIED OPTICS | 1990年 / 29卷 / 26期
关键词
D O I
10.1364/AO.29.003775
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We have constructed a correlation microscope based on the Mirau interferometer configuration using a thin silicon nitride film beam splitter. This microscope provides the amplitude and phase information for the reflected signal from a sample located on the microscope-object plane. The device is remarkably insensitive to vibrations and is self-correcting for spherical and chromatic range aberrations of the objective. An imaging theory for the correlation microscope has been derived, which predicts accurately both the transverse resolution at a sharp edge and the range resolution for a perfect plane reflector. The range resolution is slightly better than that for a scanning optical microscope using a lens with the same aperture. © 1990 Optical Society of America.
引用
收藏
页码:3775 / 3783
页数:9
相关论文
共 8 条
  • [1] CHIM SC, 1990, UNPUB REV SCI IN MAR, V61
  • [2] DAVIDSON M, 1987, P SOC PHOTOOPT INSTR, V775
  • [3] Goodman J., 2005, INTRO FOURIER OPTICS
  • [4] HOBBS PCD, 1990, J MICROSC
  • [5] KINO GS, 1990, SCANNING OPTICAL MIC
  • [6] REINHOLDSTEN PA, IN PRESS J OPT SOC A, V7
  • [7] Wilson T, 1984, THEORY PRACTICE SCAN
  • [8] REAL-TIME CONFOCAL SCANNING OPTICAL MICROSCOPE
    XIAO, GQ
    CORLE, TR
    KINO, GS
    [J]. APPLIED PHYSICS LETTERS, 1988, 53 (08) : 716 - 718