CALIBRATION OF RADIATION DETECTORS IN THE VUV USING A WALL-STABILIZED AR-CASCADE ARC

被引:12
作者
KAASE, H [1 ]
STEPHAN, HK [1 ]
机构
[1] MAX PLANCK INST PHYS & ASTROPHYS,INST EXTRATERR PHYS,D-8046 GARCHING,FED REP GER
来源
APPLIED OPTICS | 1979年 / 18卷 / 13期
关键词
D O I
10.1364/AO.18.002275
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
A VUV calibration facility is described. By means of this equipment, measurement of the responsivity of radiation detectors in the spectral range from 90 nm to 250 nm is possible. The measuring method makes use of a wall-stabilized Ar-cascade arc as a standard radiator, and by measuring the spectral transmittance of the optical arrangement, the monochromatic radiant flux, which reaches the detector or the detector system, can be given. The spectrum of the Ar arc in particular is investigated in the λ < 125-nm spectral range. © 1979 Optical Society of America.
引用
收藏
页码:2275 / 2279
页数:5
相关论文
共 28 条
  • [1] BISCHOFF K, 1968, OPTIK, V28, P183
  • [2] BISCHOFF K, 1978, PTB BER OPT, V6, P1
  • [3] ULTRAVIOLET PHOTOMETRY FROM ORBITING ASTRONOMICAL OBSERVATORY .21. ABSOLUTE ENERGY-DISTRIBUTION OF STARS IN ULTRAVIOLET
    BLESS, RC
    CODE, AD
    FAIRCHILD, ET
    [J]. ASTROPHYSICAL JOURNAL, 1976, 203 (02) : 410 - 416
  • [4] BOLDT G, 1967, Z ANGEW PHYSIK, V22, P429
  • [5] BOLDT G, 1961, 5 P INT C ION PHEN G, V1, P925
  • [6] BOLDT G, 1969, MPI BERICHT PAE EXTR, V29, P1
  • [7] VACUUM ULTRAVIOLET RADIOMETRY .3. ARGON MINI-ARC AS A NEW SECONDARY STANDARD OF SPECTRAL RADIANCE
    BRIDGES, JM
    OTT, WR
    [J]. APPLIED OPTICS, 1977, 16 (02): : 367 - 376
  • [8] CANFIELD LR, 1973, APPL OPT, V12, P1161
  • [9] MEASUREMENTS OF OPERATING-TIME VARIATION OF SPECTRAL RADIANCE OF DEUTERIUM LAMPS
    EINFELD, D
    STUCK, D
    [J]. OPTICS COMMUNICATIONS, 1976, 19 (02) : 297 - 301
  • [10] EINFELD D, 1977, SR7703 DESY REP